Endress+Hauser Liquiline CM44x

Multichannel transmitter for pH, conductivity, oxygen and other Memosens sensors, with up to 16 measured values.

The Endress+Hauser Liquiline CM44x is a multichannel transmitter that measures pH, conductivity, dissolved oxygen and other liquid values with up to eight Memosens sensors, in water treatment and in irrigation and drain water. This Modbus RTU template reads the 16 assignable measured values with status and unit, diagnostics and sensor data, and lets you write controller setpoints and digital and analog values from ModbusCloud.

Protocol
Modbus RTU
Connection
RS485, 19200 baud
Registers
827, read and write
Key readings
Measured value per AI slot · Value status · Unit of each value · Current output values · Diagnostic code · NE107 status signal · Sensor operating hours · Last calibration
Available commands
Controller setpoints · PID parameters · Digital values to the device · Analog values to the device · Modbus and Ethernet settings
Compatible models
Liquiline CM442 · Liquiline CM444 · Liquiline CM448 · Liquiline CM44Rx · Liquiline CM44P
Last updated

Registers (827)

827 read, 105 write

NameUnitAccessAddress (Hidden until you enter your email)Function (hidden)Type (hidden)
AI1 ValueRead
AI1 StatusRead
AI1 Unit CodeRead
AI2 ValueRead
AI2 StatusRead
AI2 Unit CodeRead
AI3 ValueRead
AI3 StatusRead
AI3 Unit CodeRead
AI4 ValueRead
AI4 StatusRead
AI4 Unit CodeRead
AI5 ValueRead
AI5 StatusRead
AI5 Unit CodeRead
AI6 ValueRead
AI6 StatusRead
AI6 Unit CodeRead
AI7 ValueRead
AI7 StatusRead
AI7 Unit CodeRead
AI8 ValueRead
AI8 StatusRead
AI8 Unit CodeRead
AI9 ValueRead
AI9 StatusRead
AI9 Unit CodeRead
AI10 ValueRead
AI10 StatusRead
AI10 Unit CodeRead
AI11 ValueRead
AI11 StatusRead
AI11 Unit CodeRead
AI12 ValueRead
AI12 StatusRead
AI12 Unit CodeRead
AI13 ValueRead
AI13 StatusRead
AI13 Unit CodeRead
AI14 ValueRead
AI14 StatusRead
AI14 Unit CodeRead
AI15 ValueRead
AI15 StatusRead
AI15 Unit CodeRead
AI16 ValueRead
AI16 StatusRead
AI16 Unit CodeRead
Current Output 1 Process ValueRead
Current Output 1 StatusRead
Current Output 1 Unit CodeRead
Current Output 2 Process ValueRead
Current Output 2 StatusRead
Current Output 2 Unit CodeRead
Current Output 3 Process ValueRead
Current Output 3 StatusRead
Current Output 3 Unit CodeRead
Current Output 4 Process ValueRead
Current Output 4 StatusRead
Current Output 4 Unit CodeRead
Current Output 5 Process ValueRead
Current Output 5 StatusRead
Current Output 5 Unit CodeRead
Current Output 6 Process ValueRead
Current Output 6 StatusRead
Current Output 6 Unit CodeRead
Current Output 7 Process ValueRead
Current Output 7 StatusRead
Current Output 7 Unit CodeRead
Current Output 8 Process ValueRead
Current Output 8 StatusRead
Current Output 8 Unit CodeRead
Current Output 1 CurrentRead
Current Output 2 CurrentRead
Current Output 3 CurrentRead
Current Output 4 CurrentRead
Current Output 5 CurrentRead
Current Output 6 CurrentRead
Current Output 7 CurrentRead
Current Output 8 CurrentRead
DI1 ValueRead
DI1 StatusRead
DI2 ValueRead
DI2 StatusRead
DI3 ValueRead
DI3 StatusRead
DI4 ValueRead
DI4 StatusRead
DI5 ValueRead
DI5 StatusRead
DI6 ValueRead
DI6 StatusRead
DI7 ValueRead
DI7 StatusRead
DI8 ValueRead
DI8 StatusRead
Diagnostic Status BitsRead
Highest Priority Diagnostic CodeRead
Highest Priority Diagnostic ChannelRead
Highest Priority Diagnostic StatusRead
Device Diagnostic ModuleRead
Device Diagnostic PortRead
Device Diagnostic NE107 StatusRead
Device Diagnostic CodeRead
Device Diagnostic TimestampRead
Additional Device DiagnosticsRead
Past Device Diagnostic ComponentRead
Past Device Diagnostic ModuleRead
Past Device Diagnostic PortRead
Past Device Diagnostic NE107 StatusRead
Past Device Diagnostic CodeRead
Past Device Diagnostic TimestampRead
Cleaning 1 StateRead
Cleaning 2 StateRead
Cleaning 3 StateRead
Cleaning 4 StateRead
Heartbeat Device StatusRead
Heartbeat Device Health%Read
Heartbeat Device Maintenance TimerRead
Heartbeat Device IndexRead
Device Availability%Read
Device Operating TimeRead
Device Time in FailureRead
Device Failure CountRead
Device MTBFRead
Device MTTRRead
DO1 ValueRead and write
DO1 StatusRead and write
DO2 ValueRead and write
DO2 StatusRead and write
DO3 ValueRead and write
DO3 StatusRead and write
DO4 ValueRead and write
DO4 StatusRead and write
DO5 ValueRead and write
DO5 StatusRead and write
DO6 ValueRead and write
DO6 StatusRead and write
DO7 ValueRead and write
DO7 StatusRead and write
DO8 ValueRead and write
DO8 StatusRead and write
AO1 ValueRead and write
AO1 StatusRead and write
AO1 Unit CodeRead and write
AO2 ValueRead and write
AO2 StatusRead and write
AO2 Unit CodeRead and write
AO3 ValueRead and write
AO3 StatusRead and write
AO3 Unit CodeRead and write
AO4 ValueRead and write
AO4 StatusRead and write
AO4 Unit CodeRead and write
Controller 0 SetpointRead and write
Controller 0 Proportional BandRead and write
Controller 0 Reset TimesRead and write
Controller 0 Rate TimesRead and write
Controller 1 SetpointRead and write
Controller 1 Proportional BandRead and write
Controller 1 Reset TimesRead and write
Controller 1 Rate TimesRead and write
Controller 2 SetpointRead and write
Controller 2 Proportional BandRead and write
Controller 2 Reset TimesRead and write
Controller 2 Rate TimesRead and write
Controller 3 SetpointRead and write
Controller 3 Proportional BandRead and write
Controller 3 Reset TimesRead and write
Controller 3 Rate TimesRead and write
Temperature UnitRead and write
RS485 Baud RateRead and write
RS485 Address ModeRead and write
RS485 ParityRead and write
RS485 EnableRead and write
Modbus ModeRead and write
Modbus WatchdogsRead and write
Modbus Statistics CommandRead and write
Ethernet EnableRead and write
Ethernet IP AddressRead and write
Ethernet NetmaskRead and write
Ethernet Default GatewayRead and write
Ethernet Link SettingsRead and write
Ethernet Statistics CommandRead and write
DHCP EnableRead and write
Web Server EnableRead and write
Device ResetRead and write
Modbus Byte OrderRead and write
Device Tag SettingRead and write
Clock SettingRead and write
Control Panel LockRead and write
Modbus Valid FramesRead
Modbus Invalid FramesRead
Modbus Framing ErrorsRead
Modbus CRC ErrorsRead
Modbus Parity ErrorsRead
Modbus Exception 01 CountRead
Modbus Exception 02 CountRead
Modbus Exception 03 CountRead
Modbus Exception 04 CountRead
Ethernet Received FramesRead
Ethernet Sent FramesRead
Ethernet Valid FramesRead
Ethernet Invalid FramesRead
Ethernet CRC ErrorsRead
Bus AddressRead
Device TagRead
Device IDRead
Software VersionRead
Bootloader VersionRead
Device Serial NumberRead
Display Serial NumberRead
Backplane Serial NumberRead
Power Supply Serial NumberRead
Base Module Serial NumberRead
Extension Module 1 Serial NumberRead
Extension Module 2 Serial NumberRead
Extension Module 3 Serial NumberRead
Extension Module 4 Serial NumberRead
Extension Module 5 Serial NumberRead
Extension Module 6 Serial NumberRead
Extension Module 7 Serial NumberRead
Extension Module 8 Serial NumberRead
Sensor 1 Serial NumberRead
Sensor 2 Serial NumberRead
Sensor 3 Serial NumberRead
Sensor 4 Serial NumberRead
Sensor 5 Serial NumberRead
Sensor 6 Serial NumberRead
Sensor 7 Serial NumberRead
Sensor 8 Serial NumberRead
FXAB1 Module Serial NumberRead
RS485 TerminationRead
Ethernet MAC AddressRead
Device Date and TimeRead
Sensor 1 Channel EnableRead
Sensor 1 Selected TypeRead
Sensor 1 Connected TypeRead
Sensor 1 TagRead
Sensor 1 Serial Number (Sensor Block)Read
Sensor 1 Hardware VersionRead
Sensor 1 Software VersionRead
Sensor 1 ManufacturerRead
Sensor 1 Order CodeRead
Sensor 1 Manufacturing DateRead
Sensor 1 Initial Operation DateRead
Sensor 1 Min Measured ValueRead
Sensor 1 Min Temperature°CRead
Sensor 1 Max Measured ValueRead
Sensor 1 Max Temperature°CRead
Sensor 1 Diagnostic ComponentRead
Sensor 1 Diagnostic ModuleRead
Sensor 1 Diagnostic PortRead
Sensor 1 Diagnostic NE107 StatusRead
Sensor 1 Diagnostic CodeRead
Sensor 1 Diagnostic TimestampRead
Sensor 1 Additional DiagnosticsRead
Sensor 1 Past Diagnostic ComponentRead
Sensor 1 Past Diagnostic ModuleRead
Sensor 1 Past Diagnostic PortRead
Sensor 1 Past Diagnostic NE107 StatusRead
Sensor 1 Past Diagnostic CodeRead
Sensor 1 Past Diagnostic TimestampRead
Sensor 1 Operating HourssRead
Sensor 1 Time Above Max TemperaturesRead
Sensor 1 Time Below Min TemperaturesRead
Sensor 1 Lamp Operating TimesRead
Sensor 1 Sterilization CyclesRead
Sensor 1 CIP CyclesRead
Sensor 1 Cap CalibrationsRead
Sensor 1 Cap Calibration TimersRead
Sensor 1 Cap SterilizationsRead
Sensor 1 Filter ChangesRead
Sensor 1 CalibrationsRead
Sensor 1 Calibrations Parameter 2Read
Sensor 1 Calibrations Parameter 3Read
Sensor 1 Calibrations Parameter 4Read
Sensor 1 Last CalibrationRead
Sensor 1 Last Calibration Parameter 2Read
Sensor 1 Last Calibration Parameter 3Read
Sensor 1 Last Calibration Parameter 4Read
Sensor 1 Calibration MethodRead
Sensor 1 Calibration Method Parameter 2Read
Sensor 1 Calibration Method Parameter 3Read
Sensor 1 Calibration Method Parameter 4Read
Sensor 1 Zero Point CalibrationsRead
Sensor 1 Last Zero Point CalibrationRead
Sensor 1 Zero Point Calibration MethodRead
Sensor 1 Temperature CalibrationsRead
Sensor 1 Last Temperature CalibrationRead
Sensor 1 Temperature Calibration MethodRead
Sensor 1 Heartbeat Status%Read
Sensor 1 Heartbeat Health%Read
Sensor 1 Maintenance TimerRead
Sensor 1 Heartbeat Index%Read
Sensor 1 Availability%Read
Sensor 1 Total Operating TimeRead
Sensor 1 Time in FailureRead
Sensor 1 Failure CountRead
Sensor 1 MTBFRead
Sensor 1 MTTRRead
Sensor 1 Time in CalibrationRead
Sensor 1 Calibrations and AdjustmentsRead
Sensor 1 MTBCRead
Sensor 1 SAC DatasetRead and write
Sensor 1 Nitrate DatasetRead and write
Sensor 1 Turbidity DatasetRead and write
Sensor 1 Photometer Dataset MCRead and write
Sensor 1 Photometer Dataset RCRead and write
Sensor 2 Channel EnableRead
Sensor 2 Selected TypeRead
Sensor 2 Connected TypeRead
Sensor 2 TagRead
Sensor 2 Serial Number (Sensor Block)Read
Sensor 2 Hardware VersionRead
Sensor 2 Software VersionRead
Sensor 2 ManufacturerRead
Sensor 2 Order CodeRead
Sensor 2 Manufacturing DateRead
Sensor 2 Initial Operation DateRead
Sensor 2 Min Measured ValueRead
Sensor 2 Min Temperature°CRead
Sensor 2 Max Measured ValueRead
Sensor 2 Max Temperature°CRead
Sensor 2 Diagnostic ComponentRead
Sensor 2 Diagnostic ModuleRead
Sensor 2 Diagnostic PortRead
Sensor 2 Diagnostic NE107 StatusRead
Sensor 2 Diagnostic CodeRead
Sensor 2 Diagnostic TimestampRead
Sensor 2 Additional DiagnosticsRead
Sensor 2 Past Diagnostic ComponentRead
Sensor 2 Past Diagnostic ModuleRead
Sensor 2 Past Diagnostic PortRead
Sensor 2 Past Diagnostic NE107 StatusRead
Sensor 2 Past Diagnostic CodeRead
Sensor 2 Past Diagnostic TimestampRead
Sensor 2 Operating HourssRead
Sensor 2 Time Above Max TemperaturesRead
Sensor 2 Time Below Min TemperaturesRead
Sensor 2 Lamp Operating TimesRead
Sensor 2 Sterilization CyclesRead
Sensor 2 CIP CyclesRead
Sensor 2 Cap CalibrationsRead
Sensor 2 Cap Calibration TimersRead
Sensor 2 Cap SterilizationsRead
Sensor 2 Filter ChangesRead
Sensor 2 CalibrationsRead
Sensor 2 Calibrations Parameter 2Read
Sensor 2 Calibrations Parameter 3Read
Sensor 2 Calibrations Parameter 4Read
Sensor 2 Last CalibrationRead
Sensor 2 Last Calibration Parameter 2Read
Sensor 2 Last Calibration Parameter 3Read
Sensor 2 Last Calibration Parameter 4Read
Sensor 2 Calibration MethodRead
Sensor 2 Calibration Method Parameter 2Read
Sensor 2 Calibration Method Parameter 3Read
Sensor 2 Calibration Method Parameter 4Read
Sensor 2 Zero Point CalibrationsRead
Sensor 2 Last Zero Point CalibrationRead
Sensor 2 Zero Point Calibration MethodRead
Sensor 2 Temperature CalibrationsRead
Sensor 2 Last Temperature CalibrationRead
Sensor 2 Temperature Calibration MethodRead
Sensor 2 Heartbeat Status%Read
Sensor 2 Heartbeat Health%Read
Sensor 2 Maintenance TimerRead
Sensor 2 Heartbeat Index%Read
Sensor 2 Availability%Read
Sensor 2 Total Operating TimeRead
Sensor 2 Time in FailureRead
Sensor 2 Failure CountRead
Sensor 2 MTBFRead
Sensor 2 MTTRRead
Sensor 2 Time in CalibrationRead
Sensor 2 Calibrations and AdjustmentsRead
Sensor 2 MTBCRead
Sensor 2 SAC DatasetRead and write
Sensor 2 Nitrate DatasetRead and write
Sensor 2 Turbidity DatasetRead and write
Sensor 2 Photometer Dataset MCRead and write
Sensor 2 Photometer Dataset RCRead and write
Sensor 3 Channel EnableRead
Sensor 3 Selected TypeRead
Sensor 3 Connected TypeRead
Sensor 3 TagRead
Sensor 3 Serial Number (Sensor Block)Read
Sensor 3 Hardware VersionRead
Sensor 3 Software VersionRead
Sensor 3 ManufacturerRead
Sensor 3 Order CodeRead
Sensor 3 Manufacturing DateRead
Sensor 3 Initial Operation DateRead
Sensor 3 Min Measured ValueRead
Sensor 3 Min Temperature°CRead
Sensor 3 Max Measured ValueRead
Sensor 3 Max Temperature°CRead
Sensor 3 Diagnostic ComponentRead
Sensor 3 Diagnostic ModuleRead
Sensor 3 Diagnostic PortRead
Sensor 3 Diagnostic NE107 StatusRead
Sensor 3 Diagnostic CodeRead
Sensor 3 Diagnostic TimestampRead
Sensor 3 Additional DiagnosticsRead
Sensor 3 Past Diagnostic ComponentRead
Sensor 3 Past Diagnostic ModuleRead
Sensor 3 Past Diagnostic PortRead
Sensor 3 Past Diagnostic NE107 StatusRead
Sensor 3 Past Diagnostic CodeRead
Sensor 3 Past Diagnostic TimestampRead
Sensor 3 Operating HourssRead
Sensor 3 Time Above Max TemperaturesRead
Sensor 3 Time Below Min TemperaturesRead
Sensor 3 Lamp Operating TimesRead
Sensor 3 Sterilization CyclesRead
Sensor 3 CIP CyclesRead
Sensor 3 Cap CalibrationsRead
Sensor 3 Cap Calibration TimersRead
Sensor 3 Cap SterilizationsRead
Sensor 3 Filter ChangesRead
Sensor 3 CalibrationsRead
Sensor 3 Calibrations Parameter 2Read
Sensor 3 Calibrations Parameter 3Read
Sensor 3 Calibrations Parameter 4Read
Sensor 3 Last CalibrationRead
Sensor 3 Last Calibration Parameter 2Read
Sensor 3 Last Calibration Parameter 3Read
Sensor 3 Last Calibration Parameter 4Read
Sensor 3 Calibration MethodRead
Sensor 3 Calibration Method Parameter 2Read
Sensor 3 Calibration Method Parameter 3Read
Sensor 3 Calibration Method Parameter 4Read
Sensor 3 Zero Point CalibrationsRead
Sensor 3 Last Zero Point CalibrationRead
Sensor 3 Zero Point Calibration MethodRead
Sensor 3 Temperature CalibrationsRead
Sensor 3 Last Temperature CalibrationRead
Sensor 3 Temperature Calibration MethodRead
Sensor 3 Heartbeat Status%Read
Sensor 3 Heartbeat Health%Read
Sensor 3 Maintenance TimerRead
Sensor 3 Heartbeat Index%Read
Sensor 3 Availability%Read
Sensor 3 Total Operating TimeRead
Sensor 3 Time in FailureRead
Sensor 3 Failure CountRead
Sensor 3 MTBFRead
Sensor 3 MTTRRead
Sensor 3 Time in CalibrationRead
Sensor 3 Calibrations and AdjustmentsRead
Sensor 3 MTBCRead
Sensor 3 SAC DatasetRead and write
Sensor 3 Nitrate DatasetRead and write
Sensor 3 Turbidity DatasetRead and write
Sensor 3 Photometer Dataset MCRead and write
Sensor 3 Photometer Dataset RCRead and write
Sensor 4 Channel EnableRead
Sensor 4 Selected TypeRead
Sensor 4 Connected TypeRead
Sensor 4 TagRead
Sensor 4 Serial Number (Sensor Block)Read
Sensor 4 Hardware VersionRead
Sensor 4 Software VersionRead
Sensor 4 ManufacturerRead
Sensor 4 Order CodeRead
Sensor 4 Manufacturing DateRead
Sensor 4 Initial Operation DateRead
Sensor 4 Min Measured ValueRead
Sensor 4 Min Temperature°CRead
Sensor 4 Max Measured ValueRead
Sensor 4 Max Temperature°CRead
Sensor 4 Diagnostic ComponentRead
Sensor 4 Diagnostic ModuleRead
Sensor 4 Diagnostic PortRead
Sensor 4 Diagnostic NE107 StatusRead
Sensor 4 Diagnostic CodeRead
Sensor 4 Diagnostic TimestampRead
Sensor 4 Additional DiagnosticsRead
Sensor 4 Past Diagnostic ComponentRead
Sensor 4 Past Diagnostic ModuleRead
Sensor 4 Past Diagnostic PortRead
Sensor 4 Past Diagnostic NE107 StatusRead
Sensor 4 Past Diagnostic CodeRead
Sensor 4 Past Diagnostic TimestampRead
Sensor 4 Operating HourssRead
Sensor 4 Time Above Max TemperaturesRead
Sensor 4 Time Below Min TemperaturesRead
Sensor 4 Lamp Operating TimesRead
Sensor 4 Sterilization CyclesRead
Sensor 4 CIP CyclesRead
Sensor 4 Cap CalibrationsRead
Sensor 4 Cap Calibration TimersRead
Sensor 4 Cap SterilizationsRead
Sensor 4 Filter ChangesRead
Sensor 4 CalibrationsRead
Sensor 4 Calibrations Parameter 2Read
Sensor 4 Calibrations Parameter 3Read
Sensor 4 Calibrations Parameter 4Read
Sensor 4 Last CalibrationRead
Sensor 4 Last Calibration Parameter 2Read
Sensor 4 Last Calibration Parameter 3Read
Sensor 4 Last Calibration Parameter 4Read
Sensor 4 Calibration MethodRead
Sensor 4 Calibration Method Parameter 2Read
Sensor 4 Calibration Method Parameter 3Read
Sensor 4 Calibration Method Parameter 4Read
Sensor 4 Zero Point CalibrationsRead
Sensor 4 Last Zero Point CalibrationRead
Sensor 4 Zero Point Calibration MethodRead
Sensor 4 Temperature CalibrationsRead
Sensor 4 Last Temperature CalibrationRead
Sensor 4 Temperature Calibration MethodRead
Sensor 4 Heartbeat Status%Read
Sensor 4 Heartbeat Health%Read
Sensor 4 Maintenance TimerRead
Sensor 4 Heartbeat Index%Read
Sensor 4 Availability%Read
Sensor 4 Total Operating TimeRead
Sensor 4 Time in FailureRead
Sensor 4 Failure CountRead
Sensor 4 MTBFRead
Sensor 4 MTTRRead
Sensor 4 Time in CalibrationRead
Sensor 4 Calibrations and AdjustmentsRead
Sensor 4 MTBCRead
Sensor 4 SAC DatasetRead and write
Sensor 4 Nitrate DatasetRead and write
Sensor 4 Turbidity DatasetRead and write
Sensor 4 Photometer Dataset MCRead and write
Sensor 4 Photometer Dataset RCRead and write
Sensor 5 Channel EnableRead
Sensor 5 Selected TypeRead
Sensor 5 Connected TypeRead
Sensor 5 TagRead
Sensor 5 Serial Number (Sensor Block)Read
Sensor 5 Hardware VersionRead
Sensor 5 Software VersionRead
Sensor 5 ManufacturerRead
Sensor 5 Order CodeRead
Sensor 5 Manufacturing DateRead
Sensor 5 Initial Operation DateRead
Sensor 5 Min Measured ValueRead
Sensor 5 Min Temperature°CRead
Sensor 5 Max Measured ValueRead
Sensor 5 Max Temperature°CRead
Sensor 5 Diagnostic ComponentRead
Sensor 5 Diagnostic ModuleRead
Sensor 5 Diagnostic PortRead
Sensor 5 Diagnostic NE107 StatusRead
Sensor 5 Diagnostic CodeRead
Sensor 5 Diagnostic TimestampRead
Sensor 5 Additional DiagnosticsRead
Sensor 5 Past Diagnostic ComponentRead
Sensor 5 Past Diagnostic ModuleRead
Sensor 5 Past Diagnostic PortRead
Sensor 5 Past Diagnostic NE107 StatusRead
Sensor 5 Past Diagnostic CodeRead
Sensor 5 Past Diagnostic TimestampRead
Sensor 5 Operating HourssRead
Sensor 5 Time Above Max TemperaturesRead
Sensor 5 Time Below Min TemperaturesRead
Sensor 5 Lamp Operating TimesRead
Sensor 5 Sterilization CyclesRead
Sensor 5 CIP CyclesRead
Sensor 5 Cap CalibrationsRead
Sensor 5 Cap Calibration TimersRead
Sensor 5 Cap SterilizationsRead
Sensor 5 Filter ChangesRead
Sensor 5 CalibrationsRead
Sensor 5 Calibrations Parameter 2Read
Sensor 5 Calibrations Parameter 3Read
Sensor 5 Calibrations Parameter 4Read
Sensor 5 Last CalibrationRead
Sensor 5 Last Calibration Parameter 2Read
Sensor 5 Last Calibration Parameter 3Read
Sensor 5 Last Calibration Parameter 4Read
Sensor 5 Calibration MethodRead
Sensor 5 Calibration Method Parameter 2Read
Sensor 5 Calibration Method Parameter 3Read
Sensor 5 Calibration Method Parameter 4Read
Sensor 5 Zero Point CalibrationsRead
Sensor 5 Last Zero Point CalibrationRead
Sensor 5 Zero Point Calibration MethodRead
Sensor 5 Temperature CalibrationsRead
Sensor 5 Last Temperature CalibrationRead
Sensor 5 Temperature Calibration MethodRead
Sensor 5 Heartbeat Status%Read
Sensor 5 Heartbeat Health%Read
Sensor 5 Maintenance TimerRead
Sensor 5 Heartbeat Index%Read
Sensor 5 Availability%Read
Sensor 5 Total Operating TimeRead
Sensor 5 Time in FailureRead
Sensor 5 Failure CountRead
Sensor 5 MTBFRead
Sensor 5 MTTRRead
Sensor 5 Time in CalibrationRead
Sensor 5 Calibrations and AdjustmentsRead
Sensor 5 MTBCRead
Sensor 5 SAC DatasetRead and write
Sensor 5 Nitrate DatasetRead and write
Sensor 5 Turbidity DatasetRead and write
Sensor 5 Photometer Dataset MCRead and write
Sensor 5 Photometer Dataset RCRead and write
Sensor 6 Channel EnableRead
Sensor 6 Selected TypeRead
Sensor 6 Connected TypeRead
Sensor 6 TagRead
Sensor 6 Serial Number (Sensor Block)Read
Sensor 6 Hardware VersionRead
Sensor 6 Software VersionRead
Sensor 6 ManufacturerRead
Sensor 6 Order CodeRead
Sensor 6 Manufacturing DateRead
Sensor 6 Initial Operation DateRead
Sensor 6 Min Measured ValueRead
Sensor 6 Min Temperature°CRead
Sensor 6 Max Measured ValueRead
Sensor 6 Max Temperature°CRead
Sensor 6 Diagnostic ComponentRead
Sensor 6 Diagnostic ModuleRead
Sensor 6 Diagnostic PortRead
Sensor 6 Diagnostic NE107 StatusRead
Sensor 6 Diagnostic CodeRead
Sensor 6 Diagnostic TimestampRead
Sensor 6 Additional DiagnosticsRead
Sensor 6 Past Diagnostic ComponentRead
Sensor 6 Past Diagnostic ModuleRead
Sensor 6 Past Diagnostic PortRead
Sensor 6 Past Diagnostic NE107 StatusRead
Sensor 6 Past Diagnostic CodeRead
Sensor 6 Past Diagnostic TimestampRead
Sensor 6 Operating HourssRead
Sensor 6 Time Above Max TemperaturesRead
Sensor 6 Time Below Min TemperaturesRead
Sensor 6 Lamp Operating TimesRead
Sensor 6 Sterilization CyclesRead
Sensor 6 CIP CyclesRead
Sensor 6 Cap CalibrationsRead
Sensor 6 Cap Calibration TimersRead
Sensor 6 Cap SterilizationsRead
Sensor 6 Filter ChangesRead
Sensor 6 CalibrationsRead
Sensor 6 Calibrations Parameter 2Read
Sensor 6 Calibrations Parameter 3Read
Sensor 6 Calibrations Parameter 4Read
Sensor 6 Last CalibrationRead
Sensor 6 Last Calibration Parameter 2Read
Sensor 6 Last Calibration Parameter 3Read
Sensor 6 Last Calibration Parameter 4Read
Sensor 6 Calibration MethodRead
Sensor 6 Calibration Method Parameter 2Read
Sensor 6 Calibration Method Parameter 3Read
Sensor 6 Calibration Method Parameter 4Read
Sensor 6 Zero Point CalibrationsRead
Sensor 6 Last Zero Point CalibrationRead
Sensor 6 Zero Point Calibration MethodRead
Sensor 6 Temperature CalibrationsRead
Sensor 6 Last Temperature CalibrationRead
Sensor 6 Temperature Calibration MethodRead
Sensor 6 Heartbeat Status%Read
Sensor 6 Heartbeat Health%Read
Sensor 6 Maintenance TimerRead
Sensor 6 Heartbeat Index%Read
Sensor 6 Availability%Read
Sensor 6 Total Operating TimeRead
Sensor 6 Time in FailureRead
Sensor 6 Failure CountRead
Sensor 6 MTBFRead
Sensor 6 MTTRRead
Sensor 6 Time in CalibrationRead
Sensor 6 Calibrations and AdjustmentsRead
Sensor 6 MTBCRead
Sensor 6 SAC DatasetRead and write
Sensor 6 Nitrate DatasetRead and write
Sensor 6 Turbidity DatasetRead and write
Sensor 6 Photometer Dataset MCRead and write
Sensor 6 Photometer Dataset RCRead and write
Sensor 7 Channel EnableRead
Sensor 7 Selected TypeRead
Sensor 7 Connected TypeRead
Sensor 7 TagRead
Sensor 7 Serial Number (Sensor Block)Read
Sensor 7 Hardware VersionRead
Sensor 7 Software VersionRead
Sensor 7 ManufacturerRead
Sensor 7 Order CodeRead
Sensor 7 Manufacturing DateRead
Sensor 7 Initial Operation DateRead
Sensor 7 Min Measured ValueRead
Sensor 7 Min Temperature°CRead
Sensor 7 Max Measured ValueRead
Sensor 7 Max Temperature°CRead
Sensor 7 Diagnostic ComponentRead
Sensor 7 Diagnostic ModuleRead
Sensor 7 Diagnostic PortRead
Sensor 7 Diagnostic NE107 StatusRead
Sensor 7 Diagnostic CodeRead
Sensor 7 Diagnostic TimestampRead
Sensor 7 Additional DiagnosticsRead
Sensor 7 Past Diagnostic ComponentRead
Sensor 7 Past Diagnostic ModuleRead
Sensor 7 Past Diagnostic PortRead
Sensor 7 Past Diagnostic NE107 StatusRead
Sensor 7 Past Diagnostic CodeRead
Sensor 7 Past Diagnostic TimestampRead
Sensor 7 Operating HourssRead
Sensor 7 Time Above Max TemperaturesRead
Sensor 7 Time Below Min TemperaturesRead
Sensor 7 Lamp Operating TimesRead
Sensor 7 Sterilization CyclesRead
Sensor 7 CIP CyclesRead
Sensor 7 Cap CalibrationsRead
Sensor 7 Cap Calibration TimersRead
Sensor 7 Cap SterilizationsRead
Sensor 7 Filter ChangesRead
Sensor 7 CalibrationsRead
Sensor 7 Calibrations Parameter 2Read
Sensor 7 Calibrations Parameter 3Read
Sensor 7 Calibrations Parameter 4Read
Sensor 7 Last CalibrationRead
Sensor 7 Last Calibration Parameter 2Read
Sensor 7 Last Calibration Parameter 3Read
Sensor 7 Last Calibration Parameter 4Read
Sensor 7 Calibration MethodRead
Sensor 7 Calibration Method Parameter 2Read
Sensor 7 Calibration Method Parameter 3Read
Sensor 7 Calibration Method Parameter 4Read
Sensor 7 Zero Point CalibrationsRead
Sensor 7 Last Zero Point CalibrationRead
Sensor 7 Zero Point Calibration MethodRead
Sensor 7 Temperature CalibrationsRead
Sensor 7 Last Temperature CalibrationRead
Sensor 7 Temperature Calibration MethodRead
Sensor 7 Heartbeat Status%Read
Sensor 7 Heartbeat Health%Read
Sensor 7 Maintenance TimerRead
Sensor 7 Heartbeat Index%Read
Sensor 7 Availability%Read
Sensor 7 Total Operating TimeRead
Sensor 7 Time in FailureRead
Sensor 7 Failure CountRead
Sensor 7 MTBFRead
Sensor 7 MTTRRead
Sensor 7 Time in CalibrationRead
Sensor 7 Calibrations and AdjustmentsRead
Sensor 7 MTBCRead
Sensor 7 SAC DatasetRead and write
Sensor 7 Nitrate DatasetRead and write
Sensor 7 Turbidity DatasetRead and write
Sensor 7 Photometer Dataset MCRead and write
Sensor 7 Photometer Dataset RCRead and write
Sensor 8 Channel EnableRead
Sensor 8 Selected TypeRead
Sensor 8 Connected TypeRead
Sensor 8 TagRead
Sensor 8 Serial Number (Sensor Block)Read
Sensor 8 Hardware VersionRead
Sensor 8 Software VersionRead
Sensor 8 ManufacturerRead
Sensor 8 Order CodeRead
Sensor 8 Manufacturing DateRead
Sensor 8 Initial Operation DateRead
Sensor 8 Min Measured ValueRead
Sensor 8 Min Temperature°CRead
Sensor 8 Max Measured ValueRead
Sensor 8 Max Temperature°CRead
Sensor 8 Diagnostic ComponentRead
Sensor 8 Diagnostic ModuleRead
Sensor 8 Diagnostic PortRead
Sensor 8 Diagnostic NE107 StatusRead
Sensor 8 Diagnostic CodeRead
Sensor 8 Diagnostic TimestampRead
Sensor 8 Additional DiagnosticsRead
Sensor 8 Past Diagnostic ComponentRead
Sensor 8 Past Diagnostic ModuleRead
Sensor 8 Past Diagnostic PortRead
Sensor 8 Past Diagnostic NE107 StatusRead
Sensor 8 Past Diagnostic CodeRead
Sensor 8 Past Diagnostic TimestampRead
Sensor 8 Operating HourssRead
Sensor 8 Time Above Max TemperaturesRead
Sensor 8 Time Below Min TemperaturesRead
Sensor 8 Lamp Operating TimesRead
Sensor 8 Sterilization CyclesRead
Sensor 8 CIP CyclesRead
Sensor 8 Cap CalibrationsRead
Sensor 8 Cap Calibration TimersRead
Sensor 8 Cap SterilizationsRead
Sensor 8 Filter ChangesRead
Sensor 8 CalibrationsRead
Sensor 8 Calibrations Parameter 2Read
Sensor 8 Calibrations Parameter 3Read
Sensor 8 Calibrations Parameter 4Read
Sensor 8 Last CalibrationRead
Sensor 8 Last Calibration Parameter 2Read
Sensor 8 Last Calibration Parameter 3Read
Sensor 8 Last Calibration Parameter 4Read
Sensor 8 Calibration MethodRead
Sensor 8 Calibration Method Parameter 2Read
Sensor 8 Calibration Method Parameter 3Read
Sensor 8 Calibration Method Parameter 4Read
Sensor 8 Zero Point CalibrationsRead
Sensor 8 Last Zero Point CalibrationRead
Sensor 8 Zero Point Calibration MethodRead
Sensor 8 Temperature CalibrationsRead
Sensor 8 Last Temperature CalibrationRead
Sensor 8 Temperature Calibration MethodRead
Sensor 8 Heartbeat Status%Read
Sensor 8 Heartbeat Health%Read
Sensor 8 Maintenance TimerRead
Sensor 8 Heartbeat Index%Read
Sensor 8 Availability%Read
Sensor 8 Total Operating TimeRead
Sensor 8 Time in FailureRead
Sensor 8 Failure CountRead
Sensor 8 MTBFRead
Sensor 8 MTTRRead
Sensor 8 Time in CalibrationRead
Sensor 8 Calibrations and AdjustmentsRead
Sensor 8 MTBCRead
Sensor 8 SAC DatasetRead and write
Sensor 8 Nitrate DatasetRead and write
Sensor 8 Turbidity DatasetRead and write
Sensor 8 Photometer Dataset MCRead and write
Sensor 8 Photometer Dataset RCRead and write

Loading registers…

Send your device list

If a device is not in the library yet, we build the template for free as soon as we have the register data.

Number of installations (optional)

Use cases

  • Show pH and EC of the nutrient solution and the drain water side by side on a dashboard.
  • Get an email when the pH leaves the band you set, or when the status of a value turns to Bad.
  • Alert on a diagnostic code or an NE107 failure signal, so a fouled or broken sensor is found quickly.
  • Keep 365 days of history per measured value to compare dosing and drain water over the season.
  • Change a controller setpoint on the Liquiline from ModbusCloud when the recipe changes.
  • Track sensor operating hours and the last calibration per sensor to plan maintenance.

Installation

  1. In the Liquiline menu Setup/Outputs/Modbus, assign a measured value to each AI slot you need, for example pH to AI1 and conductivity to AI2.

  2. Check the Modbus settings under Setup/General settings/Extended setup/Modbus: address 247, 19200 baud, even parity and byte order 1-0-3-2 by default. Set the watchdog to 0.

  3. Connect B (red) and A (green) of the RS485 module to the RS485 port of the ModbusCloud Gateway, with termination at both ends of the bus.

  4. In ModbusCloud, add a Modbus device with the same address and serial settings and import this template.

  5. Check that AI1 Status reads Good and AI1 Unit Code shows the unit you selected, for example pH or µS/cm.

Things to watch for

The Liquiline does not answer. What should I check?

Compare address, baud rate and parity with the Liquiline: 247, 19200 and even parity are the defaults. An address set with the switches on the module overrides the menu. If nothing comes back, swap A and B; the Liquiline labels B as the positive line. Turn on termination only on the devices at both ends of the bus.

Why does AI3 show the status Not assigned?

The Liquiline only sends what you assign. Status 3 (Not assigned) means no source is selected for that slot under Setup/Outputs/Modbus. Pick the sensor and the measured value there; the unit code then tells you in which unit the value arrives.

The values make no sense. Is the byte order wrong?

The template expects the factory byte order 1-0-3-2 in the Modbus settings of the Liquiline. If it was changed, the measured values arrive as nonsense. Set it back to 1-0-3-2; after that AI1 Value matches the display again.

Why does the Liquiline report diagnostic S969 Modbus watchdog?

By default the Liquiline expects a Modbus request at least every 5 seconds. ModbusCloud reads every 5 minutes by default, so the watchdog runs out between readings. Set the watchdog to 0 (off) under Setup/General settings/Extended setup/Modbus/Settings, or to a time longer than the interval you set in ModbusCloud.

Can I write setpoints and outputs from ModbusCloud?

Yes. Controller setpoints and PID parameters, DO1 to DO8, AO1 to AO4 and the interface settings are writable. Writing is off per gateway by default; request it in the portal. The Liquiline only uses a DO or AO value when its status is 0 (Good) and the value is selected as a source in the menu. Change one value at a time and read it back.

Want to see how it works?

Electrician working in an open distribution board