Endress+Hauser Liquiline CM44x Heartbeat Oxygen and Disinfection
Sensor health and calibration history for oxygen and chlorine sensors on Liquiline CM44x.
The Endress+Hauser Liquiline CM44x is a multi-channel transmitter that measures dissolved oxygen, chlorine and other water parameters in water treatment, horticulture and industry. This Modbus RTU template reads the Heartbeat Monitoring data of up to eight oxygen and eight disinfection sensors, so you can follow sensor health and calibration in ModbusCloud.
- Protocol
- Modbus RTU
- Connection
- RS485, 19200 baud
- Registers
- 832, read only
- Brand
- Endress+Hauser
- Key readings
- NAMUR NE107 status · Diagnosis code · Slope and zero point · Calibration dates · Operating hours · Cap operating hours
- Compatible models
- Liquiline CM442 · Liquiline CM444 · Liquiline CM448 · Liquiline CM44P · Liquiline CM44R
- Last updated
Registers (832)
832 read, 0 write
| Name | Unit | Access | Address (Hidden until you enter your email) | Function (hidden) | Type (hidden) |
|---|---|---|---|---|---|
| Disinfection 1 Diagnosis Time | Read | ||||
| Disinfection 1 Manufacturing Date | Read | ||||
| Disinfection 1 Initial Operation Date | Read | ||||
| Disinfection 1 Last Slope Calibration Date | Read | ||||
| Disinfection 1 Last Zero Point Calibration Date | Read | ||||
| Disinfection 1 Slope | Read | ||||
| Disinfection 1 Slope Change Since Last Calibration | Read | ||||
| Disinfection 1 Zero Point | Read | ||||
| Disinfection 1 Zero Point Change Since Last Calibration | Read | ||||
| Disinfection 1 Temperature Offset | K | Read | |||
| Disinfection 1 Electrical Charge | Read | ||||
| Disinfection 1 Min Measured Value | Read | ||||
| Disinfection 1 Min Measured Temperature | °C | Read | |||
| Disinfection 1 Max Measured Value | Read | ||||
| Disinfection 1 Max Measured Temperature | °C | Read | |||
| Disinfection 1 Operating Hours | s | Read | |||
| Disinfection 1 Operating Time Above 15 C | s | Read | |||
| Disinfection 1 Operating Time Above 30 C | s | Read | |||
| Disinfection 1 Operating Time Below -20 nA | s | Read | |||
| Disinfection 1 Operating Time Below -100 nA | s | Read | |||
| Disinfection 1 Sensor Cap Calibrations | Read | ||||
| Disinfection 1 Diagnosis Code | Read | ||||
| Disinfection 1 NAMUR NE107 Status | Read | ||||
| Disinfection 1 Selected Sensor Type | Read | ||||
| Disinfection 1 Number Of Calibrations | Read | ||||
| Disinfection 1 Number Of Sterilizations | Read | ||||
| Disinfection 1 Clean In Place Cycles | Read | ||||
| Disinfection 1 Number Of Autoclavings | Read | ||||
| Disinfection 1 Cap Sterilization Cycles | Read | ||||
| Disinfection 1 Cap Clean In Place Cycles | Read | ||||
| Disinfection 1 Cap Autoclavings | Read | ||||
| Disinfection 1 Electrolyte Capacity | Read | ||||
| Disinfection 1 Operating Hours Above T3 | s | Read | |||
| Disinfection 1 Cap Operating Hours | s | Read | |||
| Disinfection 1 Cap Operating Hours Above T1 | s | Read | |||
| Disinfection 1 Cap Operating Hours Above T2 | s | Read | |||
| Disinfection 1 Cap Operating Hours Above T3 | s | Read | |||
| Disinfection 1 Cap Operating Hours Above T4 | s | Read | |||
| Disinfection 1 Cap Operating Hours Above T5 | s | Read | |||
| Disinfection 1 Sensor Element | Read | ||||
| Disinfection 1 Temperature Threshold T1 | Read | ||||
| Disinfection 1 Temperature Threshold T2 | Read | ||||
| Disinfection 1 Temperature Threshold T3 | Read | ||||
| Disinfection 1 Current Threshold T4 | A | Read | |||
| Disinfection 1 Current Threshold T5 | A | Read | |||
| Disinfection 2 Diagnosis Time | Read | ||||
| Disinfection 2 Manufacturing Date | Read | ||||
| Disinfection 2 Initial Operation Date | Read | ||||
| Disinfection 2 Last Slope Calibration Date | Read | ||||
| Disinfection 2 Last Zero Point Calibration Date | Read | ||||
| Disinfection 2 Slope | Read | ||||
| Disinfection 2 Slope Change Since Last Calibration | Read | ||||
| Disinfection 2 Zero Point | Read | ||||
| Disinfection 2 Zero Point Change Since Last Calibration | Read | ||||
| Disinfection 2 Temperature Offset | K | Read | |||
| Disinfection 2 Electrical Charge | Read | ||||
| Disinfection 2 Min Measured Value | Read | ||||
| Disinfection 2 Min Measured Temperature | °C | Read | |||
| Disinfection 2 Max Measured Value | Read | ||||
| Disinfection 2 Max Measured Temperature | °C | Read | |||
| Disinfection 2 Operating Hours | s | Read | |||
| Disinfection 2 Operating Time Above 15 C | s | Read | |||
| Disinfection 2 Operating Time Above 30 C | s | Read | |||
| Disinfection 2 Operating Time Below -20 nA | s | Read | |||
| Disinfection 2 Operating Time Below -100 nA | s | Read | |||
| Disinfection 2 Sensor Cap Calibrations | Read | ||||
| Disinfection 2 Diagnosis Code | Read | ||||
| Disinfection 2 NAMUR NE107 Status | Read | ||||
| Disinfection 2 Selected Sensor Type | Read | ||||
| Disinfection 2 Number Of Calibrations | Read | ||||
| Disinfection 2 Number Of Sterilizations | Read | ||||
| Disinfection 2 Clean In Place Cycles | Read | ||||
| Disinfection 2 Number Of Autoclavings | Read | ||||
| Disinfection 2 Cap Sterilization Cycles | Read | ||||
| Disinfection 2 Cap Clean In Place Cycles | Read | ||||
| Disinfection 2 Cap Autoclavings | Read | ||||
| Disinfection 2 Electrolyte Capacity | Read | ||||
| Disinfection 2 Operating Hours Above T3 | s | Read | |||
| Disinfection 2 Cap Operating Hours | s | Read | |||
| Disinfection 2 Cap Operating Hours Above T1 | s | Read | |||
| Disinfection 2 Cap Operating Hours Above T2 | s | Read | |||
| Disinfection 2 Cap Operating Hours Above T3 | s | Read | |||
| Disinfection 2 Cap Operating Hours Above T4 | s | Read | |||
| Disinfection 2 Cap Operating Hours Above T5 | s | Read | |||
| Disinfection 2 Sensor Element | Read | ||||
| Disinfection 2 Temperature Threshold T1 | Read | ||||
| Disinfection 2 Temperature Threshold T2 | Read | ||||
| Disinfection 2 Temperature Threshold T3 | Read | ||||
| Disinfection 2 Current Threshold T4 | A | Read | |||
| Disinfection 2 Current Threshold T5 | A | Read | |||
| Disinfection 3 Diagnosis Time | Read | ||||
| Disinfection 3 Manufacturing Date | Read | ||||
| Disinfection 3 Initial Operation Date | Read | ||||
| Disinfection 3 Last Slope Calibration Date | Read | ||||
| Disinfection 3 Last Zero Point Calibration Date | Read | ||||
| Disinfection 3 Slope | Read | ||||
| Disinfection 3 Slope Change Since Last Calibration | Read | ||||
| Disinfection 3 Zero Point | Read | ||||
| Disinfection 3 Zero Point Change Since Last Calibration | Read | ||||
| Disinfection 3 Temperature Offset | K | Read | |||
| Disinfection 3 Electrical Charge | Read | ||||
| Disinfection 3 Min Measured Value | Read | ||||
| Disinfection 3 Min Measured Temperature | °C | Read | |||
| Disinfection 3 Max Measured Value | Read | ||||
| Disinfection 3 Max Measured Temperature | °C | Read | |||
| Disinfection 3 Operating Hours | s | Read | |||
| Disinfection 3 Operating Time Above 15 C | s | Read | |||
| Disinfection 3 Operating Time Above 30 C | s | Read | |||
| Disinfection 3 Operating Time Below -20 nA | s | Read | |||
| Disinfection 3 Operating Time Below -100 nA | s | Read | |||
| Disinfection 3 Sensor Cap Calibrations | Read | ||||
| Disinfection 3 Diagnosis Code | Read | ||||
| Disinfection 3 NAMUR NE107 Status | Read | ||||
| Disinfection 3 Selected Sensor Type | Read | ||||
| Disinfection 3 Number Of Calibrations | Read | ||||
| Disinfection 3 Number Of Sterilizations | Read | ||||
| Disinfection 3 Clean In Place Cycles | Read | ||||
| Disinfection 3 Number Of Autoclavings | Read | ||||
| Disinfection 3 Cap Sterilization Cycles | Read | ||||
| Disinfection 3 Cap Clean In Place Cycles | Read | ||||
| Disinfection 3 Cap Autoclavings | Read | ||||
| Disinfection 3 Electrolyte Capacity | Read | ||||
| Disinfection 3 Operating Hours Above T3 | s | Read | |||
| Disinfection 3 Cap Operating Hours | s | Read | |||
| Disinfection 3 Cap Operating Hours Above T1 | s | Read | |||
| Disinfection 3 Cap Operating Hours Above T2 | s | Read | |||
| Disinfection 3 Cap Operating Hours Above T3 | s | Read | |||
| Disinfection 3 Cap Operating Hours Above T4 | s | Read | |||
| Disinfection 3 Cap Operating Hours Above T5 | s | Read | |||
| Disinfection 3 Sensor Element | Read | ||||
| Disinfection 3 Temperature Threshold T1 | Read | ||||
| Disinfection 3 Temperature Threshold T2 | Read | ||||
| Disinfection 3 Temperature Threshold T3 | Read | ||||
| Disinfection 3 Current Threshold T4 | A | Read | |||
| Disinfection 3 Current Threshold T5 | A | Read | |||
| Disinfection 4 Diagnosis Time | Read | ||||
| Disinfection 4 Manufacturing Date | Read | ||||
| Disinfection 4 Initial Operation Date | Read | ||||
| Disinfection 4 Last Slope Calibration Date | Read | ||||
| Disinfection 4 Last Zero Point Calibration Date | Read | ||||
| Disinfection 4 Slope | Read | ||||
| Disinfection 4 Slope Change Since Last Calibration | Read | ||||
| Disinfection 4 Zero Point | Read | ||||
| Disinfection 4 Zero Point Change Since Last Calibration | Read | ||||
| Disinfection 4 Temperature Offset | K | Read | |||
| Disinfection 4 Electrical Charge | Read | ||||
| Disinfection 4 Min Measured Value | Read | ||||
| Disinfection 4 Min Measured Temperature | °C | Read | |||
| Disinfection 4 Max Measured Value | Read | ||||
| Disinfection 4 Max Measured Temperature | °C | Read | |||
| Disinfection 4 Operating Hours | s | Read | |||
| Disinfection 4 Operating Time Above 15 C | s | Read | |||
| Disinfection 4 Operating Time Above 30 C | s | Read | |||
| Disinfection 4 Operating Time Below -20 nA | s | Read | |||
| Disinfection 4 Operating Time Below -100 nA | s | Read | |||
| Disinfection 4 Sensor Cap Calibrations | Read | ||||
| Disinfection 4 Diagnosis Code | Read | ||||
| Disinfection 4 NAMUR NE107 Status | Read | ||||
| Disinfection 4 Selected Sensor Type | Read | ||||
| Disinfection 4 Number Of Calibrations | Read | ||||
| Disinfection 4 Number Of Sterilizations | Read | ||||
| Disinfection 4 Clean In Place Cycles | Read | ||||
| Disinfection 4 Number Of Autoclavings | Read | ||||
| Disinfection 4 Cap Sterilization Cycles | Read | ||||
| Disinfection 4 Cap Clean In Place Cycles | Read | ||||
| Disinfection 4 Cap Autoclavings | Read | ||||
| Disinfection 4 Electrolyte Capacity | Read | ||||
| Disinfection 4 Operating Hours Above T3 | s | Read | |||
| Disinfection 4 Cap Operating Hours | s | Read | |||
| Disinfection 4 Cap Operating Hours Above T1 | s | Read | |||
| Disinfection 4 Cap Operating Hours Above T2 | s | Read | |||
| Disinfection 4 Cap Operating Hours Above T3 | s | Read | |||
| Disinfection 4 Cap Operating Hours Above T4 | s | Read | |||
| Disinfection 4 Cap Operating Hours Above T5 | s | Read | |||
| Disinfection 4 Sensor Element | Read | ||||
| Disinfection 4 Temperature Threshold T1 | Read | ||||
| Disinfection 4 Temperature Threshold T2 | Read | ||||
| Disinfection 4 Temperature Threshold T3 | Read | ||||
| Disinfection 4 Current Threshold T4 | A | Read | |||
| Disinfection 4 Current Threshold T5 | A | Read | |||
| Disinfection 5 Diagnosis Time | Read | ||||
| Disinfection 5 Manufacturing Date | Read | ||||
| Disinfection 5 Initial Operation Date | Read | ||||
| Disinfection 5 Last Slope Calibration Date | Read | ||||
| Disinfection 5 Last Zero Point Calibration Date | Read | ||||
| Disinfection 5 Slope | Read | ||||
| Disinfection 5 Slope Change Since Last Calibration | Read | ||||
| Disinfection 5 Zero Point | Read | ||||
| Disinfection 5 Zero Point Change Since Last Calibration | Read | ||||
| Disinfection 5 Temperature Offset | K | Read | |||
| Disinfection 5 Electrical Charge | Read | ||||
| Disinfection 5 Min Measured Value | Read | ||||
| Disinfection 5 Min Measured Temperature | °C | Read | |||
| Disinfection 5 Max Measured Value | Read | ||||
| Disinfection 5 Max Measured Temperature | °C | Read | |||
| Disinfection 5 Operating Hours | s | Read | |||
| Disinfection 5 Operating Time Above 15 C | s | Read | |||
| Disinfection 5 Operating Time Above 30 C | s | Read | |||
| Disinfection 5 Operating Time Below -20 nA | s | Read | |||
| Disinfection 5 Operating Time Below -100 nA | s | Read | |||
| Disinfection 5 Sensor Cap Calibrations | Read | ||||
| Disinfection 5 Diagnosis Code | Read | ||||
| Disinfection 5 NAMUR NE107 Status | Read | ||||
| Disinfection 5 Selected Sensor Type | Read | ||||
| Disinfection 5 Number Of Calibrations | Read | ||||
| Disinfection 5 Number Of Sterilizations | Read | ||||
| Disinfection 5 Clean In Place Cycles | Read | ||||
| Disinfection 5 Number Of Autoclavings | Read | ||||
| Disinfection 5 Cap Sterilization Cycles | Read | ||||
| Disinfection 5 Cap Clean In Place Cycles | Read | ||||
| Disinfection 5 Cap Autoclavings | Read | ||||
| Disinfection 5 Electrolyte Capacity | Read | ||||
| Disinfection 5 Operating Hours Above T3 | s | Read | |||
| Disinfection 5 Cap Operating Hours | s | Read | |||
| Disinfection 5 Cap Operating Hours Above T1 | s | Read | |||
| Disinfection 5 Cap Operating Hours Above T2 | s | Read | |||
| Disinfection 5 Cap Operating Hours Above T3 | s | Read | |||
| Disinfection 5 Cap Operating Hours Above T4 | s | Read | |||
| Disinfection 5 Cap Operating Hours Above T5 | s | Read | |||
| Disinfection 5 Sensor Element | Read | ||||
| Disinfection 5 Temperature Threshold T1 | Read | ||||
| Disinfection 5 Temperature Threshold T2 | Read | ||||
| Disinfection 5 Temperature Threshold T3 | Read | ||||
| Disinfection 5 Current Threshold T4 | A | Read | |||
| Disinfection 5 Current Threshold T5 | A | Read | |||
| Disinfection 6 Diagnosis Time | Read | ||||
| Disinfection 6 Manufacturing Date | Read | ||||
| Disinfection 6 Initial Operation Date | Read | ||||
| Disinfection 6 Last Slope Calibration Date | Read | ||||
| Disinfection 6 Last Zero Point Calibration Date | Read | ||||
| Disinfection 6 Slope | Read | ||||
| Disinfection 6 Slope Change Since Last Calibration | Read | ||||
| Disinfection 6 Zero Point | Read | ||||
| Disinfection 6 Zero Point Change Since Last Calibration | Read | ||||
| Disinfection 6 Temperature Offset | K | Read | |||
| Disinfection 6 Electrical Charge | Read | ||||
| Disinfection 6 Min Measured Value | Read | ||||
| Disinfection 6 Min Measured Temperature | °C | Read | |||
| Disinfection 6 Max Measured Value | Read | ||||
| Disinfection 6 Max Measured Temperature | °C | Read | |||
| Disinfection 6 Operating Hours | s | Read | |||
| Disinfection 6 Operating Time Above 15 C | s | Read | |||
| Disinfection 6 Operating Time Above 30 C | s | Read | |||
| Disinfection 6 Operating Time Below -20 nA | s | Read | |||
| Disinfection 6 Operating Time Below -100 nA | s | Read | |||
| Disinfection 6 Sensor Cap Calibrations | Read | ||||
| Disinfection 6 Diagnosis Code | Read | ||||
| Disinfection 6 NAMUR NE107 Status | Read | ||||
| Disinfection 6 Selected Sensor Type | Read | ||||
| Disinfection 6 Number Of Calibrations | Read | ||||
| Disinfection 6 Number Of Sterilizations | Read | ||||
| Disinfection 6 Clean In Place Cycles | Read | ||||
| Disinfection 6 Number Of Autoclavings | Read | ||||
| Disinfection 6 Cap Sterilization Cycles | Read | ||||
| Disinfection 6 Cap Clean In Place Cycles | Read | ||||
| Disinfection 6 Cap Autoclavings | Read | ||||
| Disinfection 6 Electrolyte Capacity | Read | ||||
| Disinfection 6 Operating Hours Above T3 | s | Read | |||
| Disinfection 6 Cap Operating Hours | s | Read | |||
| Disinfection 6 Cap Operating Hours Above T1 | s | Read | |||
| Disinfection 6 Cap Operating Hours Above T2 | s | Read | |||
| Disinfection 6 Cap Operating Hours Above T3 | s | Read | |||
| Disinfection 6 Cap Operating Hours Above T4 | s | Read | |||
| Disinfection 6 Cap Operating Hours Above T5 | s | Read | |||
| Disinfection 6 Sensor Element | Read | ||||
| Disinfection 6 Temperature Threshold T1 | Read | ||||
| Disinfection 6 Temperature Threshold T2 | Read | ||||
| Disinfection 6 Temperature Threshold T3 | Read | ||||
| Disinfection 6 Current Threshold T4 | A | Read | |||
| Disinfection 6 Current Threshold T5 | A | Read | |||
| Disinfection 7 Diagnosis Time | Read | ||||
| Disinfection 7 Manufacturing Date | Read | ||||
| Disinfection 7 Initial Operation Date | Read | ||||
| Disinfection 7 Last Slope Calibration Date | Read | ||||
| Disinfection 7 Last Zero Point Calibration Date | Read | ||||
| Disinfection 7 Slope | Read | ||||
| Disinfection 7 Slope Change Since Last Calibration | Read | ||||
| Disinfection 7 Zero Point | Read | ||||
| Disinfection 7 Zero Point Change Since Last Calibration | Read | ||||
| Disinfection 7 Temperature Offset | K | Read | |||
| Disinfection 7 Electrical Charge | Read | ||||
| Disinfection 7 Min Measured Value | Read | ||||
| Disinfection 7 Min Measured Temperature | °C | Read | |||
| Disinfection 7 Max Measured Value | Read | ||||
| Disinfection 7 Max Measured Temperature | °C | Read | |||
| Disinfection 7 Operating Hours | s | Read | |||
| Disinfection 7 Operating Time Above 15 C | s | Read | |||
| Disinfection 7 Operating Time Above 30 C | s | Read | |||
| Disinfection 7 Operating Time Below -20 nA | s | Read | |||
| Disinfection 7 Operating Time Below -100 nA | s | Read | |||
| Disinfection 7 Sensor Cap Calibrations | Read | ||||
| Disinfection 7 Diagnosis Code | Read | ||||
| Disinfection 7 NAMUR NE107 Status | Read | ||||
| Disinfection 7 Selected Sensor Type | Read | ||||
| Disinfection 7 Number Of Calibrations | Read | ||||
| Disinfection 7 Number Of Sterilizations | Read | ||||
| Disinfection 7 Clean In Place Cycles | Read | ||||
| Disinfection 7 Number Of Autoclavings | Read | ||||
| Disinfection 7 Cap Sterilization Cycles | Read | ||||
| Disinfection 7 Cap Clean In Place Cycles | Read | ||||
| Disinfection 7 Cap Autoclavings | Read | ||||
| Disinfection 7 Electrolyte Capacity | Read | ||||
| Disinfection 7 Operating Hours Above T3 | s | Read | |||
| Disinfection 7 Cap Operating Hours | s | Read | |||
| Disinfection 7 Cap Operating Hours Above T1 | s | Read | |||
| Disinfection 7 Cap Operating Hours Above T2 | s | Read | |||
| Disinfection 7 Cap Operating Hours Above T3 | s | Read | |||
| Disinfection 7 Cap Operating Hours Above T4 | s | Read | |||
| Disinfection 7 Cap Operating Hours Above T5 | s | Read | |||
| Disinfection 7 Sensor Element | Read | ||||
| Disinfection 7 Temperature Threshold T1 | Read | ||||
| Disinfection 7 Temperature Threshold T2 | Read | ||||
| Disinfection 7 Temperature Threshold T3 | Read | ||||
| Disinfection 7 Current Threshold T4 | A | Read | |||
| Disinfection 7 Current Threshold T5 | A | Read | |||
| Disinfection 8 Diagnosis Time | Read | ||||
| Disinfection 8 Manufacturing Date | Read | ||||
| Disinfection 8 Initial Operation Date | Read | ||||
| Disinfection 8 Last Slope Calibration Date | Read | ||||
| Disinfection 8 Last Zero Point Calibration Date | Read | ||||
| Disinfection 8 Slope | Read | ||||
| Disinfection 8 Slope Change Since Last Calibration | Read | ||||
| Disinfection 8 Zero Point | Read | ||||
| Disinfection 8 Zero Point Change Since Last Calibration | Read | ||||
| Disinfection 8 Temperature Offset | K | Read | |||
| Disinfection 8 Electrical Charge | Read | ||||
| Disinfection 8 Min Measured Value | Read | ||||
| Disinfection 8 Min Measured Temperature | °C | Read | |||
| Disinfection 8 Max Measured Value | Read | ||||
| Disinfection 8 Max Measured Temperature | °C | Read | |||
| Disinfection 8 Operating Hours | s | Read | |||
| Disinfection 8 Operating Time Above 15 C | s | Read | |||
| Disinfection 8 Operating Time Above 30 C | s | Read | |||
| Disinfection 8 Operating Time Below -20 nA | s | Read | |||
| Disinfection 8 Operating Time Below -100 nA | s | Read | |||
| Disinfection 8 Sensor Cap Calibrations | Read | ||||
| Disinfection 8 Diagnosis Code | Read | ||||
| Disinfection 8 NAMUR NE107 Status | Read | ||||
| Disinfection 8 Selected Sensor Type | Read | ||||
| Disinfection 8 Number Of Calibrations | Read | ||||
| Disinfection 8 Number Of Sterilizations | Read | ||||
| Disinfection 8 Clean In Place Cycles | Read | ||||
| Disinfection 8 Number Of Autoclavings | Read | ||||
| Disinfection 8 Cap Sterilization Cycles | Read | ||||
| Disinfection 8 Cap Clean In Place Cycles | Read | ||||
| Disinfection 8 Cap Autoclavings | Read | ||||
| Disinfection 8 Electrolyte Capacity | Read | ||||
| Disinfection 8 Operating Hours Above T3 | s | Read | |||
| Disinfection 8 Cap Operating Hours | s | Read | |||
| Disinfection 8 Cap Operating Hours Above T1 | s | Read | |||
| Disinfection 8 Cap Operating Hours Above T2 | s | Read | |||
| Disinfection 8 Cap Operating Hours Above T3 | s | Read | |||
| Disinfection 8 Cap Operating Hours Above T4 | s | Read | |||
| Disinfection 8 Cap Operating Hours Above T5 | s | Read | |||
| Disinfection 8 Sensor Element | Read | ||||
| Disinfection 8 Temperature Threshold T1 | Read | ||||
| Disinfection 8 Temperature Threshold T2 | Read | ||||
| Disinfection 8 Temperature Threshold T3 | Read | ||||
| Disinfection 8 Current Threshold T4 | A | Read | |||
| Disinfection 8 Current Threshold T5 | A | Read | |||
| Oxygen 1 Diagnosis Time | Read | ||||
| Oxygen 1 Manufacturing Date | Read | ||||
| Oxygen 1 Initial Operation Date | Read | ||||
| Oxygen 1 Last Calibration Date | Read | ||||
| Oxygen 1 Last Zero Point Calibration Date | Read | ||||
| Oxygen 1 Slope | Read | ||||
| Oxygen 1 Slope Change Since Last Calibration | Read | ||||
| Oxygen 1 Zero Point | Read | ||||
| Oxygen 1 Offset 26 Change Since Last Calibration | Read | ||||
| Oxygen 1 Temperature Offset | K | Read | |||
| Oxygen 1 Electrical Charge | Read | ||||
| Oxygen 1 Min Measured Temperature | °C | Read | |||
| Oxygen 1 Max Measured Temperature | °C | Read | |||
| Oxygen 1 Operating Hours | s | Read | |||
| Oxygen 1 Operating Time Below 5 C | s | Read | |||
| Oxygen 1 Operating Time Counter 40 | s | Read | |||
| Oxygen 1 Operating Time Counter 42 | s | Read | |||
| Oxygen 1 Operating Time Counter 44 | s | Read | |||
| Oxygen 1 Operating Time Counter 46 | s | Read | |||
| Oxygen 1 Operating Time Counter 48 | s | Read | |||
| Oxygen 1 Operating Time Counter 50 | s | Read | |||
| Oxygen 1 Operating Time Counter 52 | s | Read | |||
| Oxygen 1 Operating Time Below 25 us | s | Read | |||
| Oxygen 1 Operating Time Above 40 us | s | Read | |||
| Oxygen 1 Sterilization Cycles | Read | ||||
| Oxygen 1 Clean In Place Cycles | Read | ||||
| Oxygen 1 Number Of Calibrations | Read | ||||
| Oxygen 1 Sensor Cap Calibrations | Read | ||||
| Oxygen 1 Diagnosis Code | Read | ||||
| Oxygen 1 NAMUR NE107 Status | Read | ||||
| Oxygen 1 Selected Sensor Type | Read | ||||
| Oxygen 1 Ksv Point At Oxygen Calibration | Read | ||||
| Oxygen 1 Ksv Zero Point Calibration | Read | ||||
| Oxygen 1 Tau0 Point At Oxygen Calibration | Read | ||||
| Oxygen 1 Tau0 Zero Point Calibration | Read | ||||
| Oxygen 1 Quality Index Point At Oxygen Cal | Read | ||||
| Oxygen 1 Quality Index Zero Point Cal | Read | ||||
| Oxygen 1 Calibration Slope Absolute | Read | ||||
| Oxygen 1 Calibration Slope Change Absolute | Read | ||||
| Oxygen 1 Fermenter Scaling Factor | Read | ||||
| Oxygen 1 Temperature Threshold T1 | Read | ||||
| Oxygen 1 Temperature Threshold T2 | Read | ||||
| Oxygen 1 Temperature Threshold T3 | Read | ||||
| Oxygen 1 Temperature Threshold T4 | Read | ||||
| Oxygen 1 Cap Operating Hours | s | Read | |||
| Oxygen 1 Cap Operating Hours Above T1 | s | Read | |||
| Oxygen 1 Cap Operating Hours Above T2 | s | Read | |||
| Oxygen 1 Cap Operating Hours Above T3 | s | Read | |||
| Oxygen 1 Cap Operating Hours Below T4 | s | Read | |||
| Oxygen 1 Number Of Autoclavings | Read | ||||
| Oxygen 1 Cap Clean In Place Cycles | Read | ||||
| Oxygen 1 Cap Autoclavings | Read | ||||
| Oxygen 1 Cap Sterilization Cycles | Read | ||||
| Oxygen 1 Fermenter Scaling Active | Read | ||||
| Oxygen 1 Last Fermenter Scaling Date | Read | ||||
| Oxygen 1 Point At Oxygen Calibrations | Read | ||||
| Oxygen 1 Zero Point Calibrations | Read | ||||
| Oxygen 1 Mode Point At Oxygen Calibration | Read | ||||
| Oxygen 1 Mode Zero Point Calibration | Read | ||||
| Oxygen 2 Diagnosis Time | Read | ||||
| Oxygen 2 Manufacturing Date | Read | ||||
| Oxygen 2 Initial Operation Date | Read | ||||
| Oxygen 2 Last Calibration Date | Read | ||||
| Oxygen 2 Last Zero Point Calibration Date | Read | ||||
| Oxygen 2 Slope | Read | ||||
| Oxygen 2 Slope Change Since Last Calibration | Read | ||||
| Oxygen 2 Zero Point | Read | ||||
| Oxygen 2 Offset 26 Change Since Last Calibration | Read | ||||
| Oxygen 2 Temperature Offset | K | Read | |||
| Oxygen 2 Electrical Charge | Read | ||||
| Oxygen 2 Min Measured Temperature | °C | Read | |||
| Oxygen 2 Max Measured Temperature | °C | Read | |||
| Oxygen 2 Operating Hours | s | Read | |||
| Oxygen 2 Operating Time Below 5 C | s | Read | |||
| Oxygen 2 Operating Time Counter 40 | s | Read | |||
| Oxygen 2 Operating Time Counter 42 | s | Read | |||
| Oxygen 2 Operating Time Counter 44 | s | Read | |||
| Oxygen 2 Operating Time Counter 46 | s | Read | |||
| Oxygen 2 Operating Time Counter 48 | s | Read | |||
| Oxygen 2 Operating Time Counter 50 | s | Read | |||
| Oxygen 2 Operating Time Counter 52 | s | Read | |||
| Oxygen 2 Operating Time Below 25 us | s | Read | |||
| Oxygen 2 Operating Time Above 40 us | s | Read | |||
| Oxygen 2 Sterilization Cycles | Read | ||||
| Oxygen 2 Clean In Place Cycles | Read | ||||
| Oxygen 2 Number Of Calibrations | Read | ||||
| Oxygen 2 Sensor Cap Calibrations | Read | ||||
| Oxygen 2 Diagnosis Code | Read | ||||
| Oxygen 2 NAMUR NE107 Status | Read | ||||
| Oxygen 2 Selected Sensor Type | Read | ||||
| Oxygen 2 Ksv Point At Oxygen Calibration | Read | ||||
| Oxygen 2 Ksv Zero Point Calibration | Read | ||||
| Oxygen 2 Tau0 Point At Oxygen Calibration | Read | ||||
| Oxygen 2 Tau0 Zero Point Calibration | Read | ||||
| Oxygen 2 Quality Index Point At Oxygen Cal | Read | ||||
| Oxygen 2 Quality Index Zero Point Cal | Read | ||||
| Oxygen 2 Calibration Slope Absolute | Read | ||||
| Oxygen 2 Calibration Slope Change Absolute | Read | ||||
| Oxygen 2 Fermenter Scaling Factor | Read | ||||
| Oxygen 2 Temperature Threshold T1 | Read | ||||
| Oxygen 2 Temperature Threshold T2 | Read | ||||
| Oxygen 2 Temperature Threshold T3 | Read | ||||
| Oxygen 2 Temperature Threshold T4 | Read | ||||
| Oxygen 2 Cap Operating Hours | s | Read | |||
| Oxygen 2 Cap Operating Hours Above T1 | s | Read | |||
| Oxygen 2 Cap Operating Hours Above T2 | s | Read | |||
| Oxygen 2 Cap Operating Hours Above T3 | s | Read | |||
| Oxygen 2 Cap Operating Hours Below T4 | s | Read | |||
| Oxygen 2 Number Of Autoclavings | Read | ||||
| Oxygen 2 Cap Clean In Place Cycles | Read | ||||
| Oxygen 2 Cap Autoclavings | Read | ||||
| Oxygen 2 Cap Sterilization Cycles | Read | ||||
| Oxygen 2 Fermenter Scaling Active | Read | ||||
| Oxygen 2 Last Fermenter Scaling Date | Read | ||||
| Oxygen 2 Point At Oxygen Calibrations | Read | ||||
| Oxygen 2 Zero Point Calibrations | Read | ||||
| Oxygen 2 Mode Point At Oxygen Calibration | Read | ||||
| Oxygen 2 Mode Zero Point Calibration | Read | ||||
| Oxygen 3 Diagnosis Time | Read | ||||
| Oxygen 3 Manufacturing Date | Read | ||||
| Oxygen 3 Initial Operation Date | Read | ||||
| Oxygen 3 Last Calibration Date | Read | ||||
| Oxygen 3 Last Zero Point Calibration Date | Read | ||||
| Oxygen 3 Slope | Read | ||||
| Oxygen 3 Slope Change Since Last Calibration | Read | ||||
| Oxygen 3 Zero Point | Read | ||||
| Oxygen 3 Offset 26 Change Since Last Calibration | Read | ||||
| Oxygen 3 Temperature Offset | K | Read | |||
| Oxygen 3 Electrical Charge | Read | ||||
| Oxygen 3 Min Measured Temperature | °C | Read | |||
| Oxygen 3 Max Measured Temperature | °C | Read | |||
| Oxygen 3 Operating Hours | s | Read | |||
| Oxygen 3 Operating Time Below 5 C | s | Read | |||
| Oxygen 3 Operating Time Counter 40 | s | Read | |||
| Oxygen 3 Operating Time Counter 42 | s | Read | |||
| Oxygen 3 Operating Time Counter 44 | s | Read | |||
| Oxygen 3 Operating Time Counter 46 | s | Read | |||
| Oxygen 3 Operating Time Counter 48 | s | Read | |||
| Oxygen 3 Operating Time Counter 50 | s | Read | |||
| Oxygen 3 Operating Time Counter 52 | s | Read | |||
| Oxygen 3 Operating Time Below 25 us | s | Read | |||
| Oxygen 3 Operating Time Above 40 us | s | Read | |||
| Oxygen 3 Sterilization Cycles | Read | ||||
| Oxygen 3 Clean In Place Cycles | Read | ||||
| Oxygen 3 Number Of Calibrations | Read | ||||
| Oxygen 3 Sensor Cap Calibrations | Read | ||||
| Oxygen 3 Diagnosis Code | Read | ||||
| Oxygen 3 NAMUR NE107 Status | Read | ||||
| Oxygen 3 Selected Sensor Type | Read | ||||
| Oxygen 3 Ksv Point At Oxygen Calibration | Read | ||||
| Oxygen 3 Ksv Zero Point Calibration | Read | ||||
| Oxygen 3 Tau0 Point At Oxygen Calibration | Read | ||||
| Oxygen 3 Tau0 Zero Point Calibration | Read | ||||
| Oxygen 3 Quality Index Point At Oxygen Cal | Read | ||||
| Oxygen 3 Quality Index Zero Point Cal | Read | ||||
| Oxygen 3 Calibration Slope Absolute | Read | ||||
| Oxygen 3 Calibration Slope Change Absolute | Read | ||||
| Oxygen 3 Fermenter Scaling Factor | Read | ||||
| Oxygen 3 Temperature Threshold T1 | Read | ||||
| Oxygen 3 Temperature Threshold T2 | Read | ||||
| Oxygen 3 Temperature Threshold T3 | Read | ||||
| Oxygen 3 Temperature Threshold T4 | Read | ||||
| Oxygen 3 Cap Operating Hours | s | Read | |||
| Oxygen 3 Cap Operating Hours Above T1 | s | Read | |||
| Oxygen 3 Cap Operating Hours Above T2 | s | Read | |||
| Oxygen 3 Cap Operating Hours Above T3 | s | Read | |||
| Oxygen 3 Cap Operating Hours Below T4 | s | Read | |||
| Oxygen 3 Number Of Autoclavings | Read | ||||
| Oxygen 3 Cap Clean In Place Cycles | Read | ||||
| Oxygen 3 Cap Autoclavings | Read | ||||
| Oxygen 3 Cap Sterilization Cycles | Read | ||||
| Oxygen 3 Fermenter Scaling Active | Read | ||||
| Oxygen 3 Last Fermenter Scaling Date | Read | ||||
| Oxygen 3 Point At Oxygen Calibrations | Read | ||||
| Oxygen 3 Zero Point Calibrations | Read | ||||
| Oxygen 3 Mode Point At Oxygen Calibration | Read | ||||
| Oxygen 3 Mode Zero Point Calibration | Read | ||||
| Oxygen 4 Diagnosis Time | Read | ||||
| Oxygen 4 Manufacturing Date | Read | ||||
| Oxygen 4 Initial Operation Date | Read | ||||
| Oxygen 4 Last Calibration Date | Read | ||||
| Oxygen 4 Last Zero Point Calibration Date | Read | ||||
| Oxygen 4 Slope | Read | ||||
| Oxygen 4 Slope Change Since Last Calibration | Read | ||||
| Oxygen 4 Zero Point | Read | ||||
| Oxygen 4 Offset 26 Change Since Last Calibration | Read | ||||
| Oxygen 4 Temperature Offset | K | Read | |||
| Oxygen 4 Electrical Charge | Read | ||||
| Oxygen 4 Min Measured Temperature | °C | Read | |||
| Oxygen 4 Max Measured Temperature | °C | Read | |||
| Oxygen 4 Operating Hours | s | Read | |||
| Oxygen 4 Operating Time Below 5 C | s | Read | |||
| Oxygen 4 Operating Time Counter 40 | s | Read | |||
| Oxygen 4 Operating Time Counter 42 | s | Read | |||
| Oxygen 4 Operating Time Counter 44 | s | Read | |||
| Oxygen 4 Operating Time Counter 46 | s | Read | |||
| Oxygen 4 Operating Time Counter 48 | s | Read | |||
| Oxygen 4 Operating Time Counter 50 | s | Read | |||
| Oxygen 4 Operating Time Counter 52 | s | Read | |||
| Oxygen 4 Operating Time Below 25 us | s | Read | |||
| Oxygen 4 Operating Time Above 40 us | s | Read | |||
| Oxygen 4 Sterilization Cycles | Read | ||||
| Oxygen 4 Clean In Place Cycles | Read | ||||
| Oxygen 4 Number Of Calibrations | Read | ||||
| Oxygen 4 Sensor Cap Calibrations | Read | ||||
| Oxygen 4 Diagnosis Code | Read | ||||
| Oxygen 4 NAMUR NE107 Status | Read | ||||
| Oxygen 4 Selected Sensor Type | Read | ||||
| Oxygen 4 Ksv Point At Oxygen Calibration | Read | ||||
| Oxygen 4 Ksv Zero Point Calibration | Read | ||||
| Oxygen 4 Tau0 Point At Oxygen Calibration | Read | ||||
| Oxygen 4 Tau0 Zero Point Calibration | Read | ||||
| Oxygen 4 Quality Index Point At Oxygen Cal | Read | ||||
| Oxygen 4 Quality Index Zero Point Cal | Read | ||||
| Oxygen 4 Calibration Slope Absolute | Read | ||||
| Oxygen 4 Calibration Slope Change Absolute | Read | ||||
| Oxygen 4 Fermenter Scaling Factor | Read | ||||
| Oxygen 4 Temperature Threshold T1 | Read | ||||
| Oxygen 4 Temperature Threshold T2 | Read | ||||
| Oxygen 4 Temperature Threshold T3 | Read | ||||
| Oxygen 4 Temperature Threshold T4 | Read | ||||
| Oxygen 4 Cap Operating Hours | s | Read | |||
| Oxygen 4 Cap Operating Hours Above T1 | s | Read | |||
| Oxygen 4 Cap Operating Hours Above T2 | s | Read | |||
| Oxygen 4 Cap Operating Hours Above T3 | s | Read | |||
| Oxygen 4 Cap Operating Hours Below T4 | s | Read | |||
| Oxygen 4 Number Of Autoclavings | Read | ||||
| Oxygen 4 Cap Clean In Place Cycles | Read | ||||
| Oxygen 4 Cap Autoclavings | Read | ||||
| Oxygen 4 Cap Sterilization Cycles | Read | ||||
| Oxygen 4 Fermenter Scaling Active | Read | ||||
| Oxygen 4 Last Fermenter Scaling Date | Read | ||||
| Oxygen 4 Point At Oxygen Calibrations | Read | ||||
| Oxygen 4 Zero Point Calibrations | Read | ||||
| Oxygen 4 Mode Point At Oxygen Calibration | Read | ||||
| Oxygen 4 Mode Zero Point Calibration | Read | ||||
| Oxygen 5 Diagnosis Time | Read | ||||
| Oxygen 5 Manufacturing Date | Read | ||||
| Oxygen 5 Initial Operation Date | Read | ||||
| Oxygen 5 Last Calibration Date | Read | ||||
| Oxygen 5 Last Zero Point Calibration Date | Read | ||||
| Oxygen 5 Slope | Read | ||||
| Oxygen 5 Slope Change Since Last Calibration | Read | ||||
| Oxygen 5 Zero Point | Read | ||||
| Oxygen 5 Offset 26 Change Since Last Calibration | Read | ||||
| Oxygen 5 Temperature Offset | K | Read | |||
| Oxygen 5 Electrical Charge | Read | ||||
| Oxygen 5 Min Measured Temperature | °C | Read | |||
| Oxygen 5 Max Measured Temperature | °C | Read | |||
| Oxygen 5 Operating Hours | s | Read | |||
| Oxygen 5 Operating Time Below 5 C | s | Read | |||
| Oxygen 5 Operating Time Counter 40 | s | Read | |||
| Oxygen 5 Operating Time Counter 42 | s | Read | |||
| Oxygen 5 Operating Time Counter 44 | s | Read | |||
| Oxygen 5 Operating Time Counter 46 | s | Read | |||
| Oxygen 5 Operating Time Counter 48 | s | Read | |||
| Oxygen 5 Operating Time Counter 50 | s | Read | |||
| Oxygen 5 Operating Time Counter 52 | s | Read | |||
| Oxygen 5 Operating Time Below 25 us | s | Read | |||
| Oxygen 5 Operating Time Above 40 us | s | Read | |||
| Oxygen 5 Sterilization Cycles | Read | ||||
| Oxygen 5 Clean In Place Cycles | Read | ||||
| Oxygen 5 Number Of Calibrations | Read | ||||
| Oxygen 5 Sensor Cap Calibrations | Read | ||||
| Oxygen 5 Diagnosis Code | Read | ||||
| Oxygen 5 NAMUR NE107 Status | Read | ||||
| Oxygen 5 Selected Sensor Type | Read | ||||
| Oxygen 5 Ksv Point At Oxygen Calibration | Read | ||||
| Oxygen 5 Ksv Zero Point Calibration | Read | ||||
| Oxygen 5 Tau0 Point At Oxygen Calibration | Read | ||||
| Oxygen 5 Tau0 Zero Point Calibration | Read | ||||
| Oxygen 5 Quality Index Point At Oxygen Cal | Read | ||||
| Oxygen 5 Quality Index Zero Point Cal | Read | ||||
| Oxygen 5 Calibration Slope Absolute | Read | ||||
| Oxygen 5 Calibration Slope Change Absolute | Read | ||||
| Oxygen 5 Fermenter Scaling Factor | Read | ||||
| Oxygen 5 Temperature Threshold T1 | Read | ||||
| Oxygen 5 Temperature Threshold T2 | Read | ||||
| Oxygen 5 Temperature Threshold T3 | Read | ||||
| Oxygen 5 Temperature Threshold T4 | Read | ||||
| Oxygen 5 Cap Operating Hours | s | Read | |||
| Oxygen 5 Cap Operating Hours Above T1 | s | Read | |||
| Oxygen 5 Cap Operating Hours Above T2 | s | Read | |||
| Oxygen 5 Cap Operating Hours Above T3 | s | Read | |||
| Oxygen 5 Cap Operating Hours Below T4 | s | Read | |||
| Oxygen 5 Number Of Autoclavings | Read | ||||
| Oxygen 5 Cap Clean In Place Cycles | Read | ||||
| Oxygen 5 Cap Autoclavings | Read | ||||
| Oxygen 5 Cap Sterilization Cycles | Read | ||||
| Oxygen 5 Fermenter Scaling Active | Read | ||||
| Oxygen 5 Last Fermenter Scaling Date | Read | ||||
| Oxygen 5 Point At Oxygen Calibrations | Read | ||||
| Oxygen 5 Zero Point Calibrations | Read | ||||
| Oxygen 5 Mode Point At Oxygen Calibration | Read | ||||
| Oxygen 5 Mode Zero Point Calibration | Read | ||||
| Oxygen 6 Diagnosis Time | Read | ||||
| Oxygen 6 Manufacturing Date | Read | ||||
| Oxygen 6 Initial Operation Date | Read | ||||
| Oxygen 6 Last Calibration Date | Read | ||||
| Oxygen 6 Last Zero Point Calibration Date | Read | ||||
| Oxygen 6 Slope | Read | ||||
| Oxygen 6 Slope Change Since Last Calibration | Read | ||||
| Oxygen 6 Zero Point | Read | ||||
| Oxygen 6 Offset 26 Change Since Last Calibration | Read | ||||
| Oxygen 6 Temperature Offset | K | Read | |||
| Oxygen 6 Electrical Charge | Read | ||||
| Oxygen 6 Min Measured Temperature | °C | Read | |||
| Oxygen 6 Max Measured Temperature | °C | Read | |||
| Oxygen 6 Operating Hours | s | Read | |||
| Oxygen 6 Operating Time Below 5 C | s | Read | |||
| Oxygen 6 Operating Time Counter 40 | s | Read | |||
| Oxygen 6 Operating Time Counter 42 | s | Read | |||
| Oxygen 6 Operating Time Counter 44 | s | Read | |||
| Oxygen 6 Operating Time Counter 46 | s | Read | |||
| Oxygen 6 Operating Time Counter 48 | s | Read | |||
| Oxygen 6 Operating Time Counter 50 | s | Read | |||
| Oxygen 6 Operating Time Counter 52 | s | Read | |||
| Oxygen 6 Operating Time Below 25 us | s | Read | |||
| Oxygen 6 Operating Time Above 40 us | s | Read | |||
| Oxygen 6 Sterilization Cycles | Read | ||||
| Oxygen 6 Clean In Place Cycles | Read | ||||
| Oxygen 6 Number Of Calibrations | Read | ||||
| Oxygen 6 Sensor Cap Calibrations | Read | ||||
| Oxygen 6 Diagnosis Code | Read | ||||
| Oxygen 6 NAMUR NE107 Status | Read | ||||
| Oxygen 6 Selected Sensor Type | Read | ||||
| Oxygen 6 Ksv Point At Oxygen Calibration | Read | ||||
| Oxygen 6 Ksv Zero Point Calibration | Read | ||||
| Oxygen 6 Tau0 Point At Oxygen Calibration | Read | ||||
| Oxygen 6 Tau0 Zero Point Calibration | Read | ||||
| Oxygen 6 Quality Index Point At Oxygen Cal | Read | ||||
| Oxygen 6 Quality Index Zero Point Cal | Read | ||||
| Oxygen 6 Calibration Slope Absolute | Read | ||||
| Oxygen 6 Calibration Slope Change Absolute | Read | ||||
| Oxygen 6 Fermenter Scaling Factor | Read | ||||
| Oxygen 6 Temperature Threshold T1 | Read | ||||
| Oxygen 6 Temperature Threshold T2 | Read | ||||
| Oxygen 6 Temperature Threshold T3 | Read | ||||
| Oxygen 6 Temperature Threshold T4 | Read | ||||
| Oxygen 6 Cap Operating Hours | s | Read | |||
| Oxygen 6 Cap Operating Hours Above T1 | s | Read | |||
| Oxygen 6 Cap Operating Hours Above T2 | s | Read | |||
| Oxygen 6 Cap Operating Hours Above T3 | s | Read | |||
| Oxygen 6 Cap Operating Hours Below T4 | s | Read | |||
| Oxygen 6 Number Of Autoclavings | Read | ||||
| Oxygen 6 Cap Clean In Place Cycles | Read | ||||
| Oxygen 6 Cap Autoclavings | Read | ||||
| Oxygen 6 Cap Sterilization Cycles | Read | ||||
| Oxygen 6 Fermenter Scaling Active | Read | ||||
| Oxygen 6 Last Fermenter Scaling Date | Read | ||||
| Oxygen 6 Point At Oxygen Calibrations | Read | ||||
| Oxygen 6 Zero Point Calibrations | Read | ||||
| Oxygen 6 Mode Point At Oxygen Calibration | Read | ||||
| Oxygen 6 Mode Zero Point Calibration | Read | ||||
| Oxygen 7 Diagnosis Time | Read | ||||
| Oxygen 7 Manufacturing Date | Read | ||||
| Oxygen 7 Initial Operation Date | Read | ||||
| Oxygen 7 Last Calibration Date | Read | ||||
| Oxygen 7 Last Zero Point Calibration Date | Read | ||||
| Oxygen 7 Slope | Read | ||||
| Oxygen 7 Slope Change Since Last Calibration | Read | ||||
| Oxygen 7 Zero Point | Read | ||||
| Oxygen 7 Offset 26 Change Since Last Calibration | Read | ||||
| Oxygen 7 Temperature Offset | K | Read | |||
| Oxygen 7 Electrical Charge | Read | ||||
| Oxygen 7 Min Measured Temperature | °C | Read | |||
| Oxygen 7 Max Measured Temperature | °C | Read | |||
| Oxygen 7 Operating Hours | s | Read | |||
| Oxygen 7 Operating Time Below 5 C | s | Read | |||
| Oxygen 7 Operating Time Counter 40 | s | Read | |||
| Oxygen 7 Operating Time Counter 42 | s | Read | |||
| Oxygen 7 Operating Time Counter 44 | s | Read | |||
| Oxygen 7 Operating Time Counter 46 | s | Read | |||
| Oxygen 7 Operating Time Counter 48 | s | Read | |||
| Oxygen 7 Operating Time Counter 50 | s | Read | |||
| Oxygen 7 Operating Time Counter 52 | s | Read | |||
| Oxygen 7 Operating Time Below 25 us | s | Read | |||
| Oxygen 7 Operating Time Above 40 us | s | Read | |||
| Oxygen 7 Sterilization Cycles | Read | ||||
| Oxygen 7 Clean In Place Cycles | Read | ||||
| Oxygen 7 Number Of Calibrations | Read | ||||
| Oxygen 7 Sensor Cap Calibrations | Read | ||||
| Oxygen 7 Diagnosis Code | Read | ||||
| Oxygen 7 NAMUR NE107 Status | Read | ||||
| Oxygen 7 Selected Sensor Type | Read | ||||
| Oxygen 7 Ksv Point At Oxygen Calibration | Read | ||||
| Oxygen 7 Ksv Zero Point Calibration | Read | ||||
| Oxygen 7 Tau0 Point At Oxygen Calibration | Read | ||||
| Oxygen 7 Tau0 Zero Point Calibration | Read | ||||
| Oxygen 7 Quality Index Point At Oxygen Cal | Read | ||||
| Oxygen 7 Quality Index Zero Point Cal | Read | ||||
| Oxygen 7 Calibration Slope Absolute | Read | ||||
| Oxygen 7 Calibration Slope Change Absolute | Read | ||||
| Oxygen 7 Fermenter Scaling Factor | Read | ||||
| Oxygen 7 Temperature Threshold T1 | Read | ||||
| Oxygen 7 Temperature Threshold T2 | Read | ||||
| Oxygen 7 Temperature Threshold T3 | Read | ||||
| Oxygen 7 Temperature Threshold T4 | Read | ||||
| Oxygen 7 Cap Operating Hours | s | Read | |||
| Oxygen 7 Cap Operating Hours Above T1 | s | Read | |||
| Oxygen 7 Cap Operating Hours Above T2 | s | Read | |||
| Oxygen 7 Cap Operating Hours Above T3 | s | Read | |||
| Oxygen 7 Cap Operating Hours Below T4 | s | Read | |||
| Oxygen 7 Number Of Autoclavings | Read | ||||
| Oxygen 7 Cap Clean In Place Cycles | Read | ||||
| Oxygen 7 Cap Autoclavings | Read | ||||
| Oxygen 7 Cap Sterilization Cycles | Read | ||||
| Oxygen 7 Fermenter Scaling Active | Read | ||||
| Oxygen 7 Last Fermenter Scaling Date | Read | ||||
| Oxygen 7 Point At Oxygen Calibrations | Read | ||||
| Oxygen 7 Zero Point Calibrations | Read | ||||
| Oxygen 7 Mode Point At Oxygen Calibration | Read | ||||
| Oxygen 7 Mode Zero Point Calibration | Read | ||||
| Oxygen 8 Diagnosis Time | Read | ||||
| Oxygen 8 Manufacturing Date | Read | ||||
| Oxygen 8 Initial Operation Date | Read | ||||
| Oxygen 8 Last Calibration Date | Read | ||||
| Oxygen 8 Last Zero Point Calibration Date | Read | ||||
| Oxygen 8 Slope | Read | ||||
| Oxygen 8 Slope Change Since Last Calibration | Read | ||||
| Oxygen 8 Zero Point | Read | ||||
| Oxygen 8 Offset 26 Change Since Last Calibration | Read | ||||
| Oxygen 8 Temperature Offset | K | Read | |||
| Oxygen 8 Electrical Charge | Read | ||||
| Oxygen 8 Min Measured Temperature | °C | Read | |||
| Oxygen 8 Max Measured Temperature | °C | Read | |||
| Oxygen 8 Operating Hours | s | Read | |||
| Oxygen 8 Operating Time Below 5 C | s | Read | |||
| Oxygen 8 Operating Time Counter 40 | s | Read | |||
| Oxygen 8 Operating Time Counter 42 | s | Read | |||
| Oxygen 8 Operating Time Counter 44 | s | Read | |||
| Oxygen 8 Operating Time Counter 46 | s | Read | |||
| Oxygen 8 Operating Time Counter 48 | s | Read | |||
| Oxygen 8 Operating Time Counter 50 | s | Read | |||
| Oxygen 8 Operating Time Counter 52 | s | Read | |||
| Oxygen 8 Operating Time Below 25 us | s | Read | |||
| Oxygen 8 Operating Time Above 40 us | s | Read | |||
| Oxygen 8 Sterilization Cycles | Read | ||||
| Oxygen 8 Clean In Place Cycles | Read | ||||
| Oxygen 8 Number Of Calibrations | Read | ||||
| Oxygen 8 Sensor Cap Calibrations | Read | ||||
| Oxygen 8 Diagnosis Code | Read | ||||
| Oxygen 8 NAMUR NE107 Status | Read | ||||
| Oxygen 8 Selected Sensor Type | Read | ||||
| Oxygen 8 Ksv Point At Oxygen Calibration | Read | ||||
| Oxygen 8 Ksv Zero Point Calibration | Read | ||||
| Oxygen 8 Tau0 Point At Oxygen Calibration | Read | ||||
| Oxygen 8 Tau0 Zero Point Calibration | Read | ||||
| Oxygen 8 Quality Index Point At Oxygen Cal | Read | ||||
| Oxygen 8 Quality Index Zero Point Cal | Read | ||||
| Oxygen 8 Calibration Slope Absolute | Read | ||||
| Oxygen 8 Calibration Slope Change Absolute | Read | ||||
| Oxygen 8 Fermenter Scaling Factor | Read | ||||
| Oxygen 8 Temperature Threshold T1 | Read | ||||
| Oxygen 8 Temperature Threshold T2 | Read | ||||
| Oxygen 8 Temperature Threshold T3 | Read | ||||
| Oxygen 8 Temperature Threshold T4 | Read | ||||
| Oxygen 8 Cap Operating Hours | s | Read | |||
| Oxygen 8 Cap Operating Hours Above T1 | s | Read | |||
| Oxygen 8 Cap Operating Hours Above T2 | s | Read | |||
| Oxygen 8 Cap Operating Hours Above T3 | s | Read | |||
| Oxygen 8 Cap Operating Hours Below T4 | s | Read | |||
| Oxygen 8 Number Of Autoclavings | Read | ||||
| Oxygen 8 Cap Clean In Place Cycles | Read | ||||
| Oxygen 8 Cap Autoclavings | Read | ||||
| Oxygen 8 Cap Sterilization Cycles | Read | ||||
| Oxygen 8 Fermenter Scaling Active | Read | ||||
| Oxygen 8 Last Fermenter Scaling Date | Read | ||||
| Oxygen 8 Point At Oxygen Calibrations | Read | ||||
| Oxygen 8 Zero Point Calibrations | Read | ||||
| Oxygen 8 Mode Point At Oxygen Calibration | Read | ||||
| Oxygen 8 Mode Zero Point Calibration | Read |
Loading registers…
Send your device list
If a device is not in the library yet, we build the template for free as soon as we have the register data.
Use cases
- Show the NAMUR status and diagnosis code of every oxygen and chlorine sensor on one dashboard.
- Get an alarm by email when a sensor reports maintenance required or failure.
- Follow slope and zero point after each calibration in the history to see a sensor drift.
- Plan cap and membrane replacement from the cap operating hours and calibration counts.
- Check when each sensor was last calibrated without walking to the transmitter.
Installation
Connect the RS485 module of the Liquiline CM44x to the RS485 port of the ModbusCloud Gateway, A to A and B to B, with termination at both bus ends.
Set bus address, 19200 baud and even parity in the Modbus menu of the transmitter, or set the address with the switch on the module.
Enter the Heartbeat Monitoring activation code in the transmitter if it is not active yet.
In ModbusCloud, add a Modbus device with the same address and serial settings, import this template and remove the channels you do not use.
Things to watch for
All Heartbeat values read 0. Why?
Heartbeat Monitoring needs an activation code. Without it the transmitter answers but returns 0 in every Heartbeat register. Also check that a sensor of the right type is connected to that channel; an empty channel returns 0 as well.
The dates show very large numbers. How do I read them?
The dates are 8-byte DATETIME values, shown here as a 64-bit number. ModbusCloud shows the raw value; use it to see when a date changes, for example after a calibration, rather than as a calendar date.
The float values look wrong. What should I check?
The template uses the default byte order 1-0-3-2 of the transmitter. If someone changed the Byte order parameter in the Modbus menu, the float values come out scrambled. Set it back to 1-0-3-2.
The transmitter does not answer at all. What should I check?
Check address, baud rate and parity on both sides; the default is address 247, 19200 baud, even parity. If the address switch on the RS485 module is set, it overrides the menu. Swap A and B if needed and check the termination.