Endress+Hauser Liquiline CM44x Heartbeat Turbidity, SAC, Nitrate and ISE

Sensor health and calibration history for turbidity, SAC, nitrate and ISE sensors on Liquiline CM44x.

The Endress+Hauser Liquiline CM44x is a multi-channel transmitter that measures turbidity, nitrate, organic load (SAC) and ions such as ammonium in water treatment, horticulture and industry. This Modbus RTU template reads the Heartbeat Monitoring data of up to eight sensors of each of these types, so you can follow sensor health and calibration in ModbusCloud.

Protocol
Modbus RTU
Connection
RS485, 19200 baud
Registers
736, read only
Key readings
NAMUR NE107 status · Diagnosis code · Calibration dates · Operating hours · Flash lamp counter · Slope and zero point per ISE electrode
Compatible models
Liquiline CM442 · Liquiline CM444 · Liquiline CM448 · Liquiline CM44P · Liquiline CM44R
Last updated

Registers (736)

736 read, 0 write

NameUnitAccessAddress (Hidden until you enter your email)Function (hidden)Type (hidden)
SAC 1 Calibration OffsetRead
SAC 1 Extended SACRead
SAC 1 Min Measured Temperature°CRead
SAC 1 Max Measured Temperature°CRead
SAC 1 Operating HourssRead
SAC 1 Filter Operating TimesRead
SAC 1 Flash Lamp CounterRead
SAC 1 Diagnosis TimeRead
SAC 1 Manufacturing DateRead
SAC 1 Initial Operation DateRead
SAC 1 Last Calibration DateRead
SAC 1 Dataset SwitchRead
SAC 1 Calibration MethodRead
SAC 1 Number Of CalibrationsRead
SAC 1 Diagnosis CodeRead
SAC 1 NAMUR NE107 StatusRead
SAC 1 Selected Sensor TypeRead
SAC 1 Raw ValueRead
SAC 1 Calibration FactorRead
SAC 2 Calibration OffsetRead
SAC 2 Extended SACRead
SAC 2 Min Measured Temperature°CRead
SAC 2 Max Measured Temperature°CRead
SAC 2 Operating HourssRead
SAC 2 Filter Operating TimesRead
SAC 2 Flash Lamp CounterRead
SAC 2 Diagnosis TimeRead
SAC 2 Manufacturing DateRead
SAC 2 Initial Operation DateRead
SAC 2 Last Calibration DateRead
SAC 2 Dataset SwitchRead
SAC 2 Calibration MethodRead
SAC 2 Number Of CalibrationsRead
SAC 2 Diagnosis CodeRead
SAC 2 NAMUR NE107 StatusRead
SAC 2 Selected Sensor TypeRead
SAC 2 Raw ValueRead
SAC 2 Calibration FactorRead
SAC 3 Calibration OffsetRead
SAC 3 Extended SACRead
SAC 3 Min Measured Temperature°CRead
SAC 3 Max Measured Temperature°CRead
SAC 3 Operating HourssRead
SAC 3 Filter Operating TimesRead
SAC 3 Flash Lamp CounterRead
SAC 3 Diagnosis TimeRead
SAC 3 Manufacturing DateRead
SAC 3 Initial Operation DateRead
SAC 3 Last Calibration DateRead
SAC 3 Dataset SwitchRead
SAC 3 Calibration MethodRead
SAC 3 Number Of CalibrationsRead
SAC 3 Diagnosis CodeRead
SAC 3 NAMUR NE107 StatusRead
SAC 3 Selected Sensor TypeRead
SAC 3 Raw ValueRead
SAC 3 Calibration FactorRead
SAC 4 Calibration OffsetRead
SAC 4 Extended SACRead
SAC 4 Min Measured Temperature°CRead
SAC 4 Max Measured Temperature°CRead
SAC 4 Operating HourssRead
SAC 4 Filter Operating TimesRead
SAC 4 Flash Lamp CounterRead
SAC 4 Diagnosis TimeRead
SAC 4 Manufacturing DateRead
SAC 4 Initial Operation DateRead
SAC 4 Last Calibration DateRead
SAC 4 Dataset SwitchRead
SAC 4 Calibration MethodRead
SAC 4 Number Of CalibrationsRead
SAC 4 Diagnosis CodeRead
SAC 4 NAMUR NE107 StatusRead
SAC 4 Selected Sensor TypeRead
SAC 4 Raw ValueRead
SAC 4 Calibration FactorRead
SAC 5 Calibration OffsetRead
SAC 5 Extended SACRead
SAC 5 Min Measured Temperature°CRead
SAC 5 Max Measured Temperature°CRead
SAC 5 Operating HourssRead
SAC 5 Filter Operating TimesRead
SAC 5 Flash Lamp CounterRead
SAC 5 Diagnosis TimeRead
SAC 5 Manufacturing DateRead
SAC 5 Initial Operation DateRead
SAC 5 Last Calibration DateRead
SAC 5 Dataset SwitchRead
SAC 5 Calibration MethodRead
SAC 5 Number Of CalibrationsRead
SAC 5 Diagnosis CodeRead
SAC 5 NAMUR NE107 StatusRead
SAC 5 Selected Sensor TypeRead
SAC 5 Raw ValueRead
SAC 5 Calibration FactorRead
SAC 6 Calibration OffsetRead
SAC 6 Extended SACRead
SAC 6 Min Measured Temperature°CRead
SAC 6 Max Measured Temperature°CRead
SAC 6 Operating HourssRead
SAC 6 Filter Operating TimesRead
SAC 6 Flash Lamp CounterRead
SAC 6 Diagnosis TimeRead
SAC 6 Manufacturing DateRead
SAC 6 Initial Operation DateRead
SAC 6 Last Calibration DateRead
SAC 6 Dataset SwitchRead
SAC 6 Calibration MethodRead
SAC 6 Number Of CalibrationsRead
SAC 6 Diagnosis CodeRead
SAC 6 NAMUR NE107 StatusRead
SAC 6 Selected Sensor TypeRead
SAC 6 Raw ValueRead
SAC 6 Calibration FactorRead
SAC 7 Calibration OffsetRead
SAC 7 Extended SACRead
SAC 7 Min Measured Temperature°CRead
SAC 7 Max Measured Temperature°CRead
SAC 7 Operating HourssRead
SAC 7 Filter Operating TimesRead
SAC 7 Flash Lamp CounterRead
SAC 7 Diagnosis TimeRead
SAC 7 Manufacturing DateRead
SAC 7 Initial Operation DateRead
SAC 7 Last Calibration DateRead
SAC 7 Dataset SwitchRead
SAC 7 Calibration MethodRead
SAC 7 Number Of CalibrationsRead
SAC 7 Diagnosis CodeRead
SAC 7 NAMUR NE107 StatusRead
SAC 7 Selected Sensor TypeRead
SAC 7 Raw ValueRead
SAC 7 Calibration FactorRead
SAC 8 Calibration OffsetRead
SAC 8 Extended SACRead
SAC 8 Min Measured Temperature°CRead
SAC 8 Max Measured Temperature°CRead
SAC 8 Operating HourssRead
SAC 8 Filter Operating TimesRead
SAC 8 Flash Lamp CounterRead
SAC 8 Diagnosis TimeRead
SAC 8 Manufacturing DateRead
SAC 8 Initial Operation DateRead
SAC 8 Last Calibration DateRead
SAC 8 Dataset SwitchRead
SAC 8 Calibration MethodRead
SAC 8 Number Of CalibrationsRead
SAC 8 Diagnosis CodeRead
SAC 8 NAMUR NE107 StatusRead
SAC 8 Selected Sensor TypeRead
SAC 8 Raw ValueRead
SAC 8 Calibration FactorRead
Turbidity 1 Calibration OffsetRead
Turbidity 1 Min Measured Temperature°CRead
Turbidity 1 Max Measured Temperature°CRead
Turbidity 1 Operating HourssRead
Turbidity 1 Operating Hours Offset 8sRead
Turbidity 1 Operating Hours Offset 10sRead
Turbidity 1 Diagnosis TimeRead
Turbidity 1 Manufacturing DateRead
Turbidity 1 Initial Operation DateRead
Turbidity 1 Last Calibration DateRead
Turbidity 1 Dataset SwitchRead
Turbidity 1 Number Of CalibrationsRead
Turbidity 1 NAMUR NE107 StatusRead
Turbidity 1 Diagnosis CodeRead
Turbidity 1 Selected Sensor TypeRead
Turbidity 1 Calibration FactorRead
Turbidity 2 Calibration OffsetRead
Turbidity 2 Min Measured Temperature°CRead
Turbidity 2 Max Measured Temperature°CRead
Turbidity 2 Operating HourssRead
Turbidity 2 Operating Hours Offset 8sRead
Turbidity 2 Operating Hours Offset 10sRead
Turbidity 2 Diagnosis TimeRead
Turbidity 2 Manufacturing DateRead
Turbidity 2 Initial Operation DateRead
Turbidity 2 Last Calibration DateRead
Turbidity 2 Dataset SwitchRead
Turbidity 2 Number Of CalibrationsRead
Turbidity 2 NAMUR NE107 StatusRead
Turbidity 2 Diagnosis CodeRead
Turbidity 2 Selected Sensor TypeRead
Turbidity 2 Calibration FactorRead
Turbidity 3 Calibration OffsetRead
Turbidity 3 Min Measured Temperature°CRead
Turbidity 3 Max Measured Temperature°CRead
Turbidity 3 Operating HourssRead
Turbidity 3 Operating Hours Offset 8sRead
Turbidity 3 Operating Hours Offset 10sRead
Turbidity 3 Diagnosis TimeRead
Turbidity 3 Manufacturing DateRead
Turbidity 3 Initial Operation DateRead
Turbidity 3 Last Calibration DateRead
Turbidity 3 Dataset SwitchRead
Turbidity 3 Number Of CalibrationsRead
Turbidity 3 NAMUR NE107 StatusRead
Turbidity 3 Diagnosis CodeRead
Turbidity 3 Selected Sensor TypeRead
Turbidity 3 Calibration FactorRead
Turbidity 4 Calibration OffsetRead
Turbidity 4 Min Measured Temperature°CRead
Turbidity 4 Max Measured Temperature°CRead
Turbidity 4 Operating HourssRead
Turbidity 4 Operating Hours Offset 8sRead
Turbidity 4 Operating Hours Offset 10sRead
Turbidity 4 Diagnosis TimeRead
Turbidity 4 Manufacturing DateRead
Turbidity 4 Initial Operation DateRead
Turbidity 4 Last Calibration DateRead
Turbidity 4 Dataset SwitchRead
Turbidity 4 Number Of CalibrationsRead
Turbidity 4 NAMUR NE107 StatusRead
Turbidity 4 Diagnosis CodeRead
Turbidity 4 Selected Sensor TypeRead
Turbidity 4 Calibration FactorRead
Turbidity 5 Calibration OffsetRead
Turbidity 5 Min Measured Temperature°CRead
Turbidity 5 Max Measured Temperature°CRead
Turbidity 5 Operating HourssRead
Turbidity 5 Operating Hours Offset 8sRead
Turbidity 5 Operating Hours Offset 10sRead
Turbidity 5 Diagnosis TimeRead
Turbidity 5 Manufacturing DateRead
Turbidity 5 Initial Operation DateRead
Turbidity 5 Last Calibration DateRead
Turbidity 5 Dataset SwitchRead
Turbidity 5 Number Of CalibrationsRead
Turbidity 5 NAMUR NE107 StatusRead
Turbidity 5 Diagnosis CodeRead
Turbidity 5 Selected Sensor TypeRead
Turbidity 5 Calibration FactorRead
Turbidity 6 Calibration OffsetRead
Turbidity 6 Min Measured Temperature°CRead
Turbidity 6 Max Measured Temperature°CRead
Turbidity 6 Operating HourssRead
Turbidity 6 Operating Hours Offset 8sRead
Turbidity 6 Operating Hours Offset 10sRead
Turbidity 6 Diagnosis TimeRead
Turbidity 6 Manufacturing DateRead
Turbidity 6 Initial Operation DateRead
Turbidity 6 Last Calibration DateRead
Turbidity 6 Dataset SwitchRead
Turbidity 6 Number Of CalibrationsRead
Turbidity 6 NAMUR NE107 StatusRead
Turbidity 6 Diagnosis CodeRead
Turbidity 6 Selected Sensor TypeRead
Turbidity 6 Calibration FactorRead
Turbidity 7 Calibration OffsetRead
Turbidity 7 Min Measured Temperature°CRead
Turbidity 7 Max Measured Temperature°CRead
Turbidity 7 Operating HourssRead
Turbidity 7 Operating Hours Offset 8sRead
Turbidity 7 Operating Hours Offset 10sRead
Turbidity 7 Diagnosis TimeRead
Turbidity 7 Manufacturing DateRead
Turbidity 7 Initial Operation DateRead
Turbidity 7 Last Calibration DateRead
Turbidity 7 Dataset SwitchRead
Turbidity 7 Number Of CalibrationsRead
Turbidity 7 NAMUR NE107 StatusRead
Turbidity 7 Diagnosis CodeRead
Turbidity 7 Selected Sensor TypeRead
Turbidity 7 Calibration FactorRead
Turbidity 8 Calibration OffsetRead
Turbidity 8 Min Measured Temperature°CRead
Turbidity 8 Max Measured Temperature°CRead
Turbidity 8 Operating HourssRead
Turbidity 8 Operating Hours Offset 8sRead
Turbidity 8 Operating Hours Offset 10sRead
Turbidity 8 Diagnosis TimeRead
Turbidity 8 Manufacturing DateRead
Turbidity 8 Initial Operation DateRead
Turbidity 8 Last Calibration DateRead
Turbidity 8 Dataset SwitchRead
Turbidity 8 Number Of CalibrationsRead
Turbidity 8 NAMUR NE107 StatusRead
Turbidity 8 Diagnosis CodeRead
Turbidity 8 Selected Sensor TypeRead
Turbidity 8 Calibration FactorRead
ISE 1 Min Measured ValueRead
ISE 1 Min Measured Temperature°CRead
ISE 1 Max Measured ValueRead
ISE 1 Max Measured Temperature°CRead
ISE 1 Slope Reference SlotRead
ISE 1 Slope Change Reference SlotRead
ISE 1 Zero Point Reference SlotRead
ISE 1 Zero Point Change Reference SlotRead
ISE 1 Slope Slot 1Read
ISE 1 Slope Change Slot 1Read
ISE 1 Zero Point Slot 1Read
ISE 1 Zero Point Change Slot 1Read
ISE 1 Slope Slot 2Read
ISE 1 Slope Change Slot 2Read
ISE 1 Zero Point Slot 2Read
ISE 1 Zero Point Change Slot 2Read
ISE 1 Slope Slot 3Read
ISE 1 Slope Change Slot 3Read
ISE 1 Zero Point Slot 3Read
ISE 1 Zero Point Change Slot 3Read
ISE 1 Operating HourssRead
ISE 1 Diagnosis TimeRead
ISE 1 Manufacturing DateRead
ISE 1 Initial Operation DateRead
ISE 1 Last Calibration Date Reference SlotRead
ISE 1 Last Calibration Date Slot 1Read
ISE 1 Last Calibration Date Slot 2Read
ISE 1 Last Calibration Date Slot 3Read
ISE 1 Electrode Type Reference SlotRead
ISE 1 Number Of Calibrations Reference SlotRead
ISE 1 Electrode Type Slot 1Read
ISE 1 Number Of Calibrations Slot 1Read
ISE 1 Electrode Type Slot 2Read
ISE 1 Number Of Calibrations Slot 2Read
ISE 1 Electrode Type Slot 3Read
ISE 1 Number Of Calibrations Slot 3Read
ISE 1 Diagnosis CodeRead
ISE 1 NAMUR NE107 StatusRead
ISE 1 Selected Sensor TypeRead
ISE 2 Min Measured ValueRead
ISE 2 Min Measured Temperature°CRead
ISE 2 Max Measured ValueRead
ISE 2 Max Measured Temperature°CRead
ISE 2 Slope Reference SlotRead
ISE 2 Slope Change Reference SlotRead
ISE 2 Zero Point Reference SlotRead
ISE 2 Zero Point Change Reference SlotRead
ISE 2 Slope Slot 1Read
ISE 2 Slope Change Slot 1Read
ISE 2 Zero Point Slot 1Read
ISE 2 Zero Point Change Slot 1Read
ISE 2 Slope Slot 2Read
ISE 2 Slope Change Slot 2Read
ISE 2 Zero Point Slot 2Read
ISE 2 Zero Point Change Slot 2Read
ISE 2 Slope Slot 3Read
ISE 2 Slope Change Slot 3Read
ISE 2 Zero Point Slot 3Read
ISE 2 Zero Point Change Slot 3Read
ISE 2 Operating HourssRead
ISE 2 Diagnosis TimeRead
ISE 2 Manufacturing DateRead
ISE 2 Initial Operation DateRead
ISE 2 Last Calibration Date Reference SlotRead
ISE 2 Last Calibration Date Slot 1Read
ISE 2 Last Calibration Date Slot 2Read
ISE 2 Last Calibration Date Slot 3Read
ISE 2 Electrode Type Reference SlotRead
ISE 2 Number Of Calibrations Reference SlotRead
ISE 2 Electrode Type Slot 1Read
ISE 2 Number Of Calibrations Slot 1Read
ISE 2 Electrode Type Slot 2Read
ISE 2 Number Of Calibrations Slot 2Read
ISE 2 Electrode Type Slot 3Read
ISE 2 Number Of Calibrations Slot 3Read
ISE 2 Diagnosis CodeRead
ISE 2 NAMUR NE107 StatusRead
ISE 2 Selected Sensor TypeRead
ISE 3 Min Measured ValueRead
ISE 3 Min Measured Temperature°CRead
ISE 3 Max Measured ValueRead
ISE 3 Max Measured Temperature°CRead
ISE 3 Slope Reference SlotRead
ISE 3 Slope Change Reference SlotRead
ISE 3 Zero Point Reference SlotRead
ISE 3 Zero Point Change Reference SlotRead
ISE 3 Slope Slot 1Read
ISE 3 Slope Change Slot 1Read
ISE 3 Zero Point Slot 1Read
ISE 3 Zero Point Change Slot 1Read
ISE 3 Slope Slot 2Read
ISE 3 Slope Change Slot 2Read
ISE 3 Zero Point Slot 2Read
ISE 3 Zero Point Change Slot 2Read
ISE 3 Slope Slot 3Read
ISE 3 Slope Change Slot 3Read
ISE 3 Zero Point Slot 3Read
ISE 3 Zero Point Change Slot 3Read
ISE 3 Operating HourssRead
ISE 3 Diagnosis TimeRead
ISE 3 Manufacturing DateRead
ISE 3 Initial Operation DateRead
ISE 3 Last Calibration Date Reference SlotRead
ISE 3 Last Calibration Date Slot 1Read
ISE 3 Last Calibration Date Slot 2Read
ISE 3 Last Calibration Date Slot 3Read
ISE 3 Electrode Type Reference SlotRead
ISE 3 Number Of Calibrations Reference SlotRead
ISE 3 Electrode Type Slot 1Read
ISE 3 Number Of Calibrations Slot 1Read
ISE 3 Electrode Type Slot 2Read
ISE 3 Number Of Calibrations Slot 2Read
ISE 3 Electrode Type Slot 3Read
ISE 3 Number Of Calibrations Slot 3Read
ISE 3 Diagnosis CodeRead
ISE 3 NAMUR NE107 StatusRead
ISE 3 Selected Sensor TypeRead
ISE 4 Min Measured ValueRead
ISE 4 Min Measured Temperature°CRead
ISE 4 Max Measured ValueRead
ISE 4 Max Measured Temperature°CRead
ISE 4 Slope Reference SlotRead
ISE 4 Slope Change Reference SlotRead
ISE 4 Zero Point Reference SlotRead
ISE 4 Zero Point Change Reference SlotRead
ISE 4 Slope Slot 1Read
ISE 4 Slope Change Slot 1Read
ISE 4 Zero Point Slot 1Read
ISE 4 Zero Point Change Slot 1Read
ISE 4 Slope Slot 2Read
ISE 4 Slope Change Slot 2Read
ISE 4 Zero Point Slot 2Read
ISE 4 Zero Point Change Slot 2Read
ISE 4 Slope Slot 3Read
ISE 4 Slope Change Slot 3Read
ISE 4 Zero Point Slot 3Read
ISE 4 Zero Point Change Slot 3Read
ISE 4 Operating HourssRead
ISE 4 Diagnosis TimeRead
ISE 4 Manufacturing DateRead
ISE 4 Initial Operation DateRead
ISE 4 Last Calibration Date Reference SlotRead
ISE 4 Last Calibration Date Slot 1Read
ISE 4 Last Calibration Date Slot 2Read
ISE 4 Last Calibration Date Slot 3Read
ISE 4 Electrode Type Reference SlotRead
ISE 4 Number Of Calibrations Reference SlotRead
ISE 4 Electrode Type Slot 1Read
ISE 4 Number Of Calibrations Slot 1Read
ISE 4 Electrode Type Slot 2Read
ISE 4 Number Of Calibrations Slot 2Read
ISE 4 Electrode Type Slot 3Read
ISE 4 Number Of Calibrations Slot 3Read
ISE 4 Diagnosis CodeRead
ISE 4 NAMUR NE107 StatusRead
ISE 4 Selected Sensor TypeRead
ISE 5 Min Measured ValueRead
ISE 5 Min Measured Temperature°CRead
ISE 5 Max Measured ValueRead
ISE 5 Max Measured Temperature°CRead
ISE 5 Slope Reference SlotRead
ISE 5 Slope Change Reference SlotRead
ISE 5 Zero Point Reference SlotRead
ISE 5 Zero Point Change Reference SlotRead
ISE 5 Slope Slot 1Read
ISE 5 Slope Change Slot 1Read
ISE 5 Zero Point Slot 1Read
ISE 5 Zero Point Change Slot 1Read
ISE 5 Slope Slot 2Read
ISE 5 Slope Change Slot 2Read
ISE 5 Zero Point Slot 2Read
ISE 5 Zero Point Change Slot 2Read
ISE 5 Slope Slot 3Read
ISE 5 Slope Change Slot 3Read
ISE 5 Zero Point Slot 3Read
ISE 5 Zero Point Change Slot 3Read
ISE 5 Operating HourssRead
ISE 5 Diagnosis TimeRead
ISE 5 Manufacturing DateRead
ISE 5 Initial Operation DateRead
ISE 5 Last Calibration Date Reference SlotRead
ISE 5 Last Calibration Date Slot 1Read
ISE 5 Last Calibration Date Slot 2Read
ISE 5 Last Calibration Date Slot 3Read
ISE 5 Electrode Type Reference SlotRead
ISE 5 Number Of Calibrations Reference SlotRead
ISE 5 Electrode Type Slot 1Read
ISE 5 Number Of Calibrations Slot 1Read
ISE 5 Electrode Type Slot 2Read
ISE 5 Number Of Calibrations Slot 2Read
ISE 5 Electrode Type Slot 3Read
ISE 5 Number Of Calibrations Slot 3Read
ISE 5 Diagnosis CodeRead
ISE 5 NAMUR NE107 StatusRead
ISE 5 Selected Sensor TypeRead
ISE 6 Min Measured ValueRead
ISE 6 Min Measured Temperature°CRead
ISE 6 Max Measured ValueRead
ISE 6 Max Measured Temperature°CRead
ISE 6 Slope Reference SlotRead
ISE 6 Slope Change Reference SlotRead
ISE 6 Zero Point Reference SlotRead
ISE 6 Zero Point Change Reference SlotRead
ISE 6 Slope Slot 1Read
ISE 6 Slope Change Slot 1Read
ISE 6 Zero Point Slot 1Read
ISE 6 Zero Point Change Slot 1Read
ISE 6 Slope Slot 2Read
ISE 6 Slope Change Slot 2Read
ISE 6 Zero Point Slot 2Read
ISE 6 Zero Point Change Slot 2Read
ISE 6 Slope Slot 3Read
ISE 6 Slope Change Slot 3Read
ISE 6 Zero Point Slot 3Read
ISE 6 Zero Point Change Slot 3Read
ISE 6 Operating HourssRead
ISE 6 Diagnosis TimeRead
ISE 6 Manufacturing DateRead
ISE 6 Initial Operation DateRead
ISE 6 Last Calibration Date Reference SlotRead
ISE 6 Last Calibration Date Slot 1Read
ISE 6 Last Calibration Date Slot 2Read
ISE 6 Last Calibration Date Slot 3Read
ISE 6 Electrode Type Reference SlotRead
ISE 6 Number Of Calibrations Reference SlotRead
ISE 6 Electrode Type Slot 1Read
ISE 6 Number Of Calibrations Slot 1Read
ISE 6 Electrode Type Slot 2Read
ISE 6 Number Of Calibrations Slot 2Read
ISE 6 Electrode Type Slot 3Read
ISE 6 Number Of Calibrations Slot 3Read
ISE 6 Diagnosis CodeRead
ISE 6 NAMUR NE107 StatusRead
ISE 6 Selected Sensor TypeRead
ISE 7 Min Measured ValueRead
ISE 7 Min Measured Temperature°CRead
ISE 7 Max Measured ValueRead
ISE 7 Max Measured Temperature°CRead
ISE 7 Slope Reference SlotRead
ISE 7 Slope Change Reference SlotRead
ISE 7 Zero Point Reference SlotRead
ISE 7 Zero Point Change Reference SlotRead
ISE 7 Slope Slot 1Read
ISE 7 Slope Change Slot 1Read
ISE 7 Zero Point Slot 1Read
ISE 7 Zero Point Change Slot 1Read
ISE 7 Slope Slot 2Read
ISE 7 Slope Change Slot 2Read
ISE 7 Zero Point Slot 2Read
ISE 7 Zero Point Change Slot 2Read
ISE 7 Slope Slot 3Read
ISE 7 Slope Change Slot 3Read
ISE 7 Zero Point Slot 3Read
ISE 7 Zero Point Change Slot 3Read
ISE 7 Operating HourssRead
ISE 7 Diagnosis TimeRead
ISE 7 Manufacturing DateRead
ISE 7 Initial Operation DateRead
ISE 7 Last Calibration Date Reference SlotRead
ISE 7 Last Calibration Date Slot 1Read
ISE 7 Last Calibration Date Slot 2Read
ISE 7 Last Calibration Date Slot 3Read
ISE 7 Electrode Type Reference SlotRead
ISE 7 Number Of Calibrations Reference SlotRead
ISE 7 Electrode Type Slot 1Read
ISE 7 Number Of Calibrations Slot 1Read
ISE 7 Electrode Type Slot 2Read
ISE 7 Number Of Calibrations Slot 2Read
ISE 7 Electrode Type Slot 3Read
ISE 7 Number Of Calibrations Slot 3Read
ISE 7 Diagnosis CodeRead
ISE 7 NAMUR NE107 StatusRead
ISE 7 Selected Sensor TypeRead
ISE 8 Min Measured ValueRead
ISE 8 Min Measured Temperature°CRead
ISE 8 Max Measured ValueRead
ISE 8 Max Measured Temperature°CRead
ISE 8 Slope Reference SlotRead
ISE 8 Slope Change Reference SlotRead
ISE 8 Zero Point Reference SlotRead
ISE 8 Zero Point Change Reference SlotRead
ISE 8 Slope Slot 1Read
ISE 8 Slope Change Slot 1Read
ISE 8 Zero Point Slot 1Read
ISE 8 Zero Point Change Slot 1Read
ISE 8 Slope Slot 2Read
ISE 8 Slope Change Slot 2Read
ISE 8 Zero Point Slot 2Read
ISE 8 Zero Point Change Slot 2Read
ISE 8 Slope Slot 3Read
ISE 8 Slope Change Slot 3Read
ISE 8 Zero Point Slot 3Read
ISE 8 Zero Point Change Slot 3Read
ISE 8 Operating HourssRead
ISE 8 Diagnosis TimeRead
ISE 8 Manufacturing DateRead
ISE 8 Initial Operation DateRead
ISE 8 Last Calibration Date Reference SlotRead
ISE 8 Last Calibration Date Slot 1Read
ISE 8 Last Calibration Date Slot 2Read
ISE 8 Last Calibration Date Slot 3Read
ISE 8 Electrode Type Reference SlotRead
ISE 8 Number Of Calibrations Reference SlotRead
ISE 8 Electrode Type Slot 1Read
ISE 8 Number Of Calibrations Slot 1Read
ISE 8 Electrode Type Slot 2Read
ISE 8 Number Of Calibrations Slot 2Read
ISE 8 Electrode Type Slot 3Read
ISE 8 Number Of Calibrations Slot 3Read
ISE 8 Diagnosis CodeRead
ISE 8 NAMUR NE107 StatusRead
ISE 8 Selected Sensor TypeRead
Nitrate 1 Calibration OffsetRead
Nitrate 1 Min Measured Temperature°CRead
Nitrate 1 Max Measured Temperature°CRead
Nitrate 1 Operating HourssRead
Nitrate 1 Filter Operating TimesRead
Nitrate 1 Flash Lamp CounterRead
Nitrate 1 Diagnosis TimeRead
Nitrate 1 Manufacturing DateRead
Nitrate 1 Initial Operation DateRead
Nitrate 1 Last Calibration DateRead
Nitrate 1 Dataset SwitchRead
Nitrate 1 Calibration MethodRead
Nitrate 1 Number Of CalibrationsRead
Nitrate 1 Diagnosis CodeRead
Nitrate 1 NAMUR NE107 StatusRead
Nitrate 1 Selected Sensor TypeRead
Nitrate 1 Raw ValueRead
Nitrate 1 Calibration FactorRead
Nitrate 2 Calibration OffsetRead
Nitrate 2 Min Measured Temperature°CRead
Nitrate 2 Max Measured Temperature°CRead
Nitrate 2 Operating HourssRead
Nitrate 2 Filter Operating TimesRead
Nitrate 2 Flash Lamp CounterRead
Nitrate 2 Diagnosis TimeRead
Nitrate 2 Manufacturing DateRead
Nitrate 2 Initial Operation DateRead
Nitrate 2 Last Calibration DateRead
Nitrate 2 Dataset SwitchRead
Nitrate 2 Calibration MethodRead
Nitrate 2 Number Of CalibrationsRead
Nitrate 2 Diagnosis CodeRead
Nitrate 2 NAMUR NE107 StatusRead
Nitrate 2 Selected Sensor TypeRead
Nitrate 2 Raw ValueRead
Nitrate 2 Calibration FactorRead
Nitrate 3 Calibration OffsetRead
Nitrate 3 Min Measured Temperature°CRead
Nitrate 3 Max Measured Temperature°CRead
Nitrate 3 Operating HourssRead
Nitrate 3 Filter Operating TimesRead
Nitrate 3 Flash Lamp CounterRead
Nitrate 3 Diagnosis TimeRead
Nitrate 3 Manufacturing DateRead
Nitrate 3 Initial Operation DateRead
Nitrate 3 Last Calibration DateRead
Nitrate 3 Dataset SwitchRead
Nitrate 3 Calibration MethodRead
Nitrate 3 Number Of CalibrationsRead
Nitrate 3 Diagnosis CodeRead
Nitrate 3 NAMUR NE107 StatusRead
Nitrate 3 Selected Sensor TypeRead
Nitrate 3 Raw ValueRead
Nitrate 3 Calibration FactorRead
Nitrate 4 Calibration OffsetRead
Nitrate 4 Min Measured Temperature°CRead
Nitrate 4 Max Measured Temperature°CRead
Nitrate 4 Operating HourssRead
Nitrate 4 Filter Operating TimesRead
Nitrate 4 Flash Lamp CounterRead
Nitrate 4 Diagnosis TimeRead
Nitrate 4 Manufacturing DateRead
Nitrate 4 Initial Operation DateRead
Nitrate 4 Last Calibration DateRead
Nitrate 4 Dataset SwitchRead
Nitrate 4 Calibration MethodRead
Nitrate 4 Number Of CalibrationsRead
Nitrate 4 Diagnosis CodeRead
Nitrate 4 NAMUR NE107 StatusRead
Nitrate 4 Selected Sensor TypeRead
Nitrate 4 Raw ValueRead
Nitrate 4 Calibration FactorRead
Nitrate 5 Calibration OffsetRead
Nitrate 5 Min Measured Temperature°CRead
Nitrate 5 Max Measured Temperature°CRead
Nitrate 5 Operating HourssRead
Nitrate 5 Filter Operating TimesRead
Nitrate 5 Flash Lamp CounterRead
Nitrate 5 Diagnosis TimeRead
Nitrate 5 Manufacturing DateRead
Nitrate 5 Initial Operation DateRead
Nitrate 5 Last Calibration DateRead
Nitrate 5 Dataset SwitchRead
Nitrate 5 Calibration MethodRead
Nitrate 5 Number Of CalibrationsRead
Nitrate 5 Diagnosis CodeRead
Nitrate 5 NAMUR NE107 StatusRead
Nitrate 5 Selected Sensor TypeRead
Nitrate 5 Raw ValueRead
Nitrate 5 Calibration FactorRead
Nitrate 6 Calibration OffsetRead
Nitrate 6 Min Measured Temperature°CRead
Nitrate 6 Max Measured Temperature°CRead
Nitrate 6 Operating HourssRead
Nitrate 6 Filter Operating TimesRead
Nitrate 6 Flash Lamp CounterRead
Nitrate 6 Diagnosis TimeRead
Nitrate 6 Manufacturing DateRead
Nitrate 6 Initial Operation DateRead
Nitrate 6 Last Calibration DateRead
Nitrate 6 Dataset SwitchRead
Nitrate 6 Calibration MethodRead
Nitrate 6 Number Of CalibrationsRead
Nitrate 6 Diagnosis CodeRead
Nitrate 6 NAMUR NE107 StatusRead
Nitrate 6 Selected Sensor TypeRead
Nitrate 6 Raw ValueRead
Nitrate 6 Calibration FactorRead
Nitrate 7 Calibration OffsetRead
Nitrate 7 Min Measured Temperature°CRead
Nitrate 7 Max Measured Temperature°CRead
Nitrate 7 Operating HourssRead
Nitrate 7 Filter Operating TimesRead
Nitrate 7 Flash Lamp CounterRead
Nitrate 7 Diagnosis TimeRead
Nitrate 7 Manufacturing DateRead
Nitrate 7 Initial Operation DateRead
Nitrate 7 Last Calibration DateRead
Nitrate 7 Dataset SwitchRead
Nitrate 7 Calibration MethodRead
Nitrate 7 Number Of CalibrationsRead
Nitrate 7 Diagnosis CodeRead
Nitrate 7 NAMUR NE107 StatusRead
Nitrate 7 Selected Sensor TypeRead
Nitrate 7 Raw ValueRead
Nitrate 7 Calibration FactorRead
Nitrate 8 Calibration OffsetRead
Nitrate 8 Min Measured Temperature°CRead
Nitrate 8 Max Measured Temperature°CRead
Nitrate 8 Operating HourssRead
Nitrate 8 Filter Operating TimesRead
Nitrate 8 Flash Lamp CounterRead
Nitrate 8 Diagnosis TimeRead
Nitrate 8 Manufacturing DateRead
Nitrate 8 Initial Operation DateRead
Nitrate 8 Last Calibration DateRead
Nitrate 8 Dataset SwitchRead
Nitrate 8 Calibration MethodRead
Nitrate 8 Number Of CalibrationsRead
Nitrate 8 Diagnosis CodeRead
Nitrate 8 NAMUR NE107 StatusRead
Nitrate 8 Selected Sensor TypeRead
Nitrate 8 Raw ValueRead
Nitrate 8 Calibration FactorRead

Loading registers…

Send your device list

If a device is not in the library yet, we build the template for free as soon as we have the register data.

Number of installations (optional)

Use cases

  • Show the NAMUR status and diagnosis code of every turbidity, SAC, nitrate and ISE sensor on one dashboard.
  • Get an alarm by email when a sensor reports maintenance required or failure.
  • Follow slope and zero point of each ISE electrode after calibration to see when an electrode wears out.
  • Plan lamp and filter service on SAC and nitrate photometers from the flash lamp counter and filter operating time.
  • Check which turbidity dataset is active and when each sensor was last calibrated.

Installation

  1. Connect the RS485 module of the Liquiline CM44x to the RS485 port of the ModbusCloud Gateway, A to A and B to B, with termination at both bus ends.

  2. Set bus address, 19200 baud and even parity in the Modbus menu of the transmitter, or set the address with the switch on the module.

  3. Enter the Heartbeat Monitoring activation code in the transmitter if it is not active yet.

  4. In ModbusCloud, add a Modbus device with the same address and serial settings, import this template and remove the channels and sensor types you do not use.

Things to watch for

All Heartbeat values read 0. Why?

Heartbeat Monitoring needs an activation code. Without it the transmitter answers but returns 0 in every Heartbeat register. Also check that a sensor of the right type is connected to that channel; an empty channel or a channel with another sensor type returns 0 as well.

The dates show very large numbers. How do I read them?

The dates are 8-byte DATETIME values, shown here as a 64-bit number. ModbusCloud shows the raw value; use it to see when a date changes, for example after a calibration, rather than as a calendar date.

The float values look wrong. What should I check?

The template uses the default byte order 1-0-3-2 of the transmitter. If someone changed the Byte order parameter in the Modbus menu, the float values come out scrambled. Set it back to 1-0-3-2.

The transmitter does not answer at all. What should I check?

Check address, baud rate and parity on both sides; the default is address 247, 19200 baud, even parity. If the address switch on the RS485 module is set, it overrides the menu. Swap A and B if needed and check the termination.

Want to see how it works?